Araújo A N, Lima J L
Department of Physical Chemistry, University of Oporto, Faculty of Pharmacy, Porto, Portugal.
J Trace Elem Electrolytes Health Dis. 1989 Jun;3(2):97-101.
This paper describes three flow injection analysis (FIA) systems for the automatic determination of sodium, potassium, lithium, calcium, magnesium, zinc, copper and iron in certain biological fluids and compares the results obtained to those of flame photometry and atomic absorption spectrometry. The set-ups were designed to allow the samples to be prepared in the same manner as that used for batch procedures with the same analytical instrument. For determinations requiring a high sample dilution, it was found of definite advantage to split the stream and pass large amounts of liquid through the detection systems. The comparison of the results obtained by FIA and conventional methods yielded correlation coefficients in the range 0.990 to 0.999. The proposed methodology show good precision, with variation coefficients between 0.5% to 5%.
本文介绍了三种流动注射分析(FIA)系统,用于自动测定某些生物流体中的钠、钾、锂、钙、镁、锌、铜和铁,并将所得结果与火焰光度法和原子吸收光谱法的结果进行比较。这些装置的设计使得样品能够以与使用相同分析仪器的批量程序相同的方式制备。对于需要高样品稀释度的测定,发现将流路分流并使大量液体通过检测系统具有明显优势。FIA法与传统方法所得结果的比较产生了0.990至0.999范围内的相关系数。所提出的方法具有良好的精密度,变异系数在0.5%至5%之间。