T1-weighted MR image contrast around a cryoablation iceball: a phantom study and initial comparison with in vivo findings.

作者信息

Overduin Christiaan G, Bomers Joyce G R, Jenniskens Sjoerd F M, Hoes Maarten F, Ten Haken Bennie, de Lange Frank, Fütterer Jurgen J, Scheenen Tom W J

机构信息

Department of Radiology and Nuclear Medicine, Radboud University Medical Centre, Nijmegen 6525GA, The Netherlands.

Department of Medical Physics, Medisch Spectrum Twente, Enschede 7500KA, The Netherlands.

出版信息

Med Phys. 2014 Nov;41(11):112301. doi: 10.1118/1.4896824.

Abstract

PURPOSE

To correlate T1-weighted (T1w) magnetic resonance (MR) image contrast around a cryoablation iceball with temperature in a phantom study and to compare this to its in vivo appearance during MR-guided focal cryoablation of prostate cancer.

METHODS

A MR-compatible cryoneedle was inserted into identical gel phantoms (n = 3) on a 1.5 T MR system. Two fiber-optic temperature sensors were placed parallel to the needle. A fast 3D T1w gradient echo (GRE) sequence (TR/TE/FA = 4.81/1.98/6°) was used to monitor iceball progression. Normalized signal intensity (SI) was correlated with temperature. The same T1w sequence was used during MR-guided prostate cryoablation in ten consecutive patients at the authors' institution. In vivo findings were quantitatively compared to the phantom data.

RESULTS

In the phantom study, the cryoablation iceball appeared in the T1w MR images as a sharply delineated signal void. A 2.2 ± 0.2 mm wide hyperintense rim directly surrounded the iceball at cooled but nonfreezing temperatures (<20 °C) in the gel. Normalized SI was maximum at 8.4 ± 2.4 °C, showing a 35.6%-43.0% (mean 40.5%) increase with respect to baseline before cooling. In the clinical procedures, the same image contrast was observed in vivo in all patients. In vivo, width of the hyperintense rim was 1.6 ± 0.6 mm. Normalized SI increases with respect to nontreated prostate ranged 28.4%-55.6% (mean 36.8%). On quantitative analysis, normalized SI changes along a linear region of interest from surrounding tissue onto the iceball center were similar between the patients and phantom setting (root mean square difference 0.06).

CONCLUSIONS

The hyperintense rim around the iceball in fast T1w GRE images corresponded to cooled but nonfreezing temperatures (<20 °C) proximal to the frozen zone. The same image contrast was observed both in a phantom study as well as in vivo in the human prostate during cryotherapy. Potentially, monitoring of this rim could be useful in order to maintain a safe margin from at-risk tissues during MR-guided prostate cryoablation procedures.

摘要

目的

在一项体模研究中,将冷冻消融冰球周围的T1加权(T1w)磁共振(MR)图像对比度与温度进行关联,并将其与前列腺癌MR引导下聚焦冷冻消融期间的体内表现进行比较。

方法

将一个与MR兼容的冷冻针插入1.5 T MR系统上的相同凝胶体模(n = 3)中。将两个光纤温度传感器与针平行放置。使用快速3D T1w梯度回波(GRE)序列(TR/TE/FA = 4.81/1.98/6°)监测冰球进展。将归一化信号强度(SI)与温度进行关联。在作者所在机构对连续10例患者进行MR引导的前列腺冷冻消融时使用相同的T1w序列。将体内发现与体模数据进行定量比较。

结果

在体模研究中,冷冻消融冰球在T1w MR图像中表现为边界清晰的信号缺失。在凝胶中冷却但未冻结的温度(<20°C)下,冰球周围直接环绕着一个宽2.2±0.2 mm的高强度边缘。归一化SI在8.4±2.4°C时最大,相对于冷却前的基线增加了35.6%-43.0%(平均40.5%)。在临床操作中,所有患者体内均观察到相同的图像对比度。在体内,高强度边缘的宽度为1.6±0.6 mm。相对于未治疗的前列腺,归一化SI增加范围为28.4%-55.6%(平均36.8%)。经定量分析,患者和体模设置中沿着从周围组织到冰球中心的线性感兴趣区域的归一化SI变化相似(均方根差异为0.06)。

结论

快速T1w GRE图像中冰球周围的高强度边缘对应于冷冻区近端冷却但未冻结的温度(<20°C)。在体模研究以及人类前列腺冷冻治疗的体内均观察到相同的图像对比度。潜在地,在MR引导的前列腺冷冻消融手术期间,监测这个边缘可能有助于与高危组织保持安全边缘。