Suppr超能文献

三个卫生系统中一氧化碳单次呼吸弥散量测定仪的准确性

Single-breath diffusing capacity for carbon monoxide instrument accuracy across 3 health systems.

作者信息

Hegewald Matthew J, Markewitz Boaz A, Wilson Emily L, Gallo Heather M, Jensen Robert L

机构信息

Intermountain Medical Center, Murray, Utah University of Utah, Salt Lake City, Utah.

University of Utah, Salt Lake City, Utah.

出版信息

Respir Care. 2015 Mar;60(3):430-6. doi: 10.4187/respcare.03512. Epub 2014 Nov 11.

Abstract

BACKGROUND

Measuring diffusing capacity of the lung for carbon monoxide (DLCO) is complex and associated with wide intra- and inter-laboratory variability. Increased D(LCO) variability may have important clinical consequences. The objective of the study was to assess instrument performance across hospital pulmonary function testing laboratories using a D(LCO) simulator that produces precise and repeatable D(LCO) values.

METHODS

D(LCO) instruments were tested with CO gas concentrations representing medium and high range D(LCO) values. The absolute difference between observed and target D(LCO) value was used to determine measurement accuracy; accuracy was defined as an average deviation from the target value of < 2.0 mL/min/mm Hg. Accuracy of inspired volume measurement and gas sensors were also determined.

RESULTS

Twenty-three instruments were tested across 3 healthcare systems. The mean absolute deviation from the target value was 1.80 mL/min/mm Hg (range 0.24-4.23) with 10 of 23 instruments (43%) being inaccurate. High volume laboratories performed better than low volume laboratories, although the difference was not significant. There was no significant difference among the instruments by manufacturers. Inspired volume was not accurate in 48% of devices; mean absolute deviation from target value was 3.7%. Instrument gas analyzers performed adequately in all instruments.

CONCLUSIONS

D(LCO) instrument accuracy was unacceptable in 43% of devices. Instrument inaccuracy can be primarily attributed to errors in inspired volume measurement and not gas analyzer performance. D(LCO) instrument performance may be improved by regular testing with a simulator. Caution should be used when comparing D(LCO) results reported from different laboratories.

摘要

背景

测量肺一氧化碳弥散量(DLCO)很复杂,且实验室内部和实验室之间的变异性很大。DLCO变异性增加可能会产生重要的临床后果。本研究的目的是使用能产生精确且可重复的DLCO值的DLCO模拟器,评估各医院肺功能检测实验室的仪器性能。

方法

使用代表中、高范围DLCO值的一氧化碳气体浓度对DLCO仪器进行测试。观察到的DLCO值与目标值之间的绝对差值用于确定测量准确性;准确性定义为与目标值的平均偏差<2.0 mL/min/mm Hg。还测定了吸气量测量和气体传感器的准确性。

结果

在3个医疗系统中对23台仪器进行了测试。与目标值的平均绝对偏差为1.80 mL/min/mm Hg(范围0.24 - 4.23),23台仪器中有10台(43%)不准确。高检测量的实验室比低检测量的实验室表现更好,尽管差异不显著。不同制造商的仪器之间没有显著差异。48%的设备吸气量不准确;与目标值的平均绝对偏差为3.7%。所有仪器中的气体分析仪性能均良好。

结论

43%的设备中DLCO仪器准确性不可接受。仪器不准确主要可归因于吸气量测量误差,而非气体分析仪性能。通过使用模拟器定期测试,可能会提高DLCO仪器的性能。比较不同实验室报告的DLCO结果时应谨慎。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验