在应变弹性成像中,以声学耦合剂为参考评估咬肌弹性。

Evaluation of the masseter muscle elasticity with the use of acoustic coupling agents as references in strain sonoelastography.

作者信息

Nakayama M, Ariji Y, Nishiyama W, Ariji E

机构信息

Department of Oral and Maxillofacial Radiology, School of Dentistry, Aichi-Gakuin University, Nagoya, Japan.

出版信息

Dentomaxillofac Radiol. 2015;44(3):20140258. doi: 10.1259/dmfr.20140258. Epub 2014 Nov 20.

Abstract

OBJECTIVES

To verify the use of a single coupling agent as a reference to obtain the elasticity index (EI) ratios and to investigate the EI ratios of the masseter muscles of healthy volunteers.

METHODS

Muscle phantoms with known elasticity (20, 40 and 60 kPa in the Young's modulus) were examined by strain-type sonoelastography using a coupling agent as the reference. Eight examiners tested soft (with 7 kPa) and hard (with 40 kpa) reference coupling agents separately. The correlation coefficients were determined between the EI ratio and Young's modulus of muscle phantoms. The interclass correlation coefficients were calculated for inter- and intraexaminer agreement.

RESULTS

Strong correlations were found between the EI ratios and Young's modulus for both soft and hard references. The variations of the EI ratios were larger with soft coupling agents than those with hard coupling agents, and they increased in phantoms with 60 kPa elasticity. There were no differences in the EI ratios of the masseter muscle at rest between males and females or between the right and left sides. The ratio increased during clenching.

CONCLUSIONS

The hard reference coupling agent was suitable for obtaining EI ratio of the masseter muscle. No differences were found in the EI ratios of the masseter muscle either between sexes or between the right and left sides at rest, and the ratios increased with the widening of their variations during clenching.

摘要

目的

验证使用单一耦合剂作为参考来获取弹性指数(EI)比值,并研究健康志愿者咬肌的EI比值。

方法

使用应变型超声弹性成像技术,以耦合剂作为参考,对已知弹性(杨氏模量分别为20、40和60kPa)的肌肉模型进行检查。八名检查者分别测试软质(7kPa)和硬质(40kPa)参考耦合剂。确定EI比值与肌肉模型杨氏模量之间的相关系数。计算检查者间和检查者内一致性的组内相关系数。

结果

软质和硬质参考的EI比值与杨氏模量之间均存在强相关性。软质耦合剂的EI比值变化比硬质耦合剂的大,且在弹性为60kPa的模型中增大。男性和女性在静息状态下咬肌的EI比值以及左右两侧之间均无差异。咬紧时该比值增加。

结论

硬质参考耦合剂适用于获取咬肌的EI比值。在静息状态下,咬肌的EI比值在性别之间或左右两侧之间均未发现差异,且在咬紧时随着其变化范围的扩大而增加。

相似文献

1
Evaluation of the masseter muscle elasticity with the use of acoustic coupling agents as references in strain sonoelastography.
Dentomaxillofac Radiol. 2015;44(3):20140258. doi: 10.1259/dmfr.20140258. Epub 2014 Nov 20.
6
Ex vivo and in vivo assessment of the non-linearity of elasticity properties of breast tissues for quantitative strain elastography.
Ultrasound Med Biol. 2014 Aug;40(8):1755-68. doi: 10.1016/j.ultrasmedbio.2014.02.005. Epub 2014 May 5.
7
Determination of Reference Values of the Masseter Muscle Stiffness in Healthy Adults Using Shear Wave Elastography.
Int J Environ Res Public Health. 2021 Sep 5;18(17):9371. doi: 10.3390/ijerph18179371.
8
Quantitative sonoelastography of the uterine cervix by interposition of a synthetic reference material.
Acta Obstet Gynecol Scand. 2013 Nov;92(11):1244-9. doi: 10.1111/aogs.12246. Epub 2013 Sep 16.
9
Quantitative assessment of temporomandibular disc and masseter muscle with shear wave elastography.
Oral Radiol. 2022 Jan;38(1):49-56. doi: 10.1007/s11282-021-00524-8. Epub 2021 Apr 5.

引用本文的文献

2
Aging-associated differences between perioral and trunk muscle characteristics.
Sci Rep. 2024 Nov 12;14(1):27745. doi: 10.1038/s41598-024-79268-4.
3
Ultrasonographic evaluation of the masseter muscle in patients with temporomandibular joint degeneration.
Imaging Sci Dent. 2023 Dec;53(4):355-363. doi: 10.5624/isd.20230134. Epub 2023 Sep 25.
5
Experimental pain and fatigue induced by excessive chewing.
BMC Oral Health. 2020 Jun 29;20(1):179. doi: 10.1186/s12903-020-01161-z.
7
Magnetic resonance and sonographic imagings of masticatory muscle myalgia in temporomandibular disorder patients.
Jpn Dent Sci Rev. 2017 Feb;53(1):11-17. doi: 10.1016/j.jdsr.2016.05.001. Epub 2016 Sep 6.
8
Sonoelastography in the musculoskeletal system: Current role and future directions.
World J Radiol. 2016 Nov 28;8(11):868-879. doi: 10.4329/wjr.v8.i11.868.

本文引用的文献

1
Ex vivo and in vivo assessment of the non-linearity of elasticity properties of breast tissues for quantitative strain elastography.
Ultrasound Med Biol. 2014 Aug;40(8):1755-68. doi: 10.1016/j.ultrasmedbio.2014.02.005. Epub 2014 May 5.
2
Effects of compression force on elasticity index and elasticity ratio in ultrasound elastography.
Dentomaxillofac Radiol. 2014;43(4):20130392. doi: 10.1259/dmfr.20130392. Epub 2014 Feb 7.
4
Measurement of muscle architecture concurrently with muscle hardness using ultrasound strain elastography.
Acta Radiol. 2014 Sep;55(7):833-9. doi: 10.1177/0284185113507565. Epub 2013 Oct 9.
5
Prediction of the muscle strength by the muscle thickness and hardness using ultrasound muscle hardness meter.
Springerplus. 2013 Sep 12;2:457. doi: 10.1186/2193-1801-2-457. eCollection 2013.
7
Effect of a 5-week static stretching program on hardness of the gastrocnemius muscle.
Scand J Med Sci Sports. 2014 Dec;24(6):950-7. doi: 10.1111/sms.12111. Epub 2013 Aug 15.
8
Acute effect of static stretching on hardness of the gastrocnemius muscle.
Med Sci Sports Exerc. 2013 Jul;45(7):1348-54. doi: 10.1249/MSS.0b013e3182850e17.
9
Ultrasound elastography for musculoskeletal applications.
Br J Radiol. 2012 Nov;85(1019):1435-45. doi: 10.1259/bjr/93042867.
10
Reliability and validity of quantifying absolute muscle hardness using ultrasound elastography.
PLoS One. 2012;7(9):e45764. doi: 10.1371/journal.pone.0045764. Epub 2012 Sep 21.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验