Nakayama M, Ariji Y, Nishiyama W, Ariji E
Department of Oral and Maxillofacial Radiology, School of Dentistry, Aichi-Gakuin University, Nagoya, Japan.
Dentomaxillofac Radiol. 2015;44(3):20140258. doi: 10.1259/dmfr.20140258. Epub 2014 Nov 20.
To verify the use of a single coupling agent as a reference to obtain the elasticity index (EI) ratios and to investigate the EI ratios of the masseter muscles of healthy volunteers.
Muscle phantoms with known elasticity (20, 40 and 60 kPa in the Young's modulus) were examined by strain-type sonoelastography using a coupling agent as the reference. Eight examiners tested soft (with 7 kPa) and hard (with 40 kpa) reference coupling agents separately. The correlation coefficients were determined between the EI ratio and Young's modulus of muscle phantoms. The interclass correlation coefficients were calculated for inter- and intraexaminer agreement.
Strong correlations were found between the EI ratios and Young's modulus for both soft and hard references. The variations of the EI ratios were larger with soft coupling agents than those with hard coupling agents, and they increased in phantoms with 60 kPa elasticity. There were no differences in the EI ratios of the masseter muscle at rest between males and females or between the right and left sides. The ratio increased during clenching.
The hard reference coupling agent was suitable for obtaining EI ratio of the masseter muscle. No differences were found in the EI ratios of the masseter muscle either between sexes or between the right and left sides at rest, and the ratios increased with the widening of their variations during clenching.
验证使用单一耦合剂作为参考来获取弹性指数(EI)比值,并研究健康志愿者咬肌的EI比值。
使用应变型超声弹性成像技术,以耦合剂作为参考,对已知弹性(杨氏模量分别为20、40和60kPa)的肌肉模型进行检查。八名检查者分别测试软质(7kPa)和硬质(40kPa)参考耦合剂。确定EI比值与肌肉模型杨氏模量之间的相关系数。计算检查者间和检查者内一致性的组内相关系数。
软质和硬质参考的EI比值与杨氏模量之间均存在强相关性。软质耦合剂的EI比值变化比硬质耦合剂的大,且在弹性为60kPa的模型中增大。男性和女性在静息状态下咬肌的EI比值以及左右两侧之间均无差异。咬紧时该比值增加。
硬质参考耦合剂适用于获取咬肌的EI比值。在静息状态下,咬肌的EI比值在性别之间或左右两侧之间均未发现差异,且在咬紧时随着其变化范围的扩大而增加。