Department of Obstetrics and Gynecology, Aarhus University Hospital, Aarhus, Denmark.
Acta Obstet Gynecol Scand. 2013 Nov;92(11):1244-9. doi: 10.1111/aogs.12246. Epub 2013 Sep 16.
To develop a reference material that allows quantitative elastography of the uterine cervix using the calculation of the approximate tissue stiffness expressed as Young's modulus (N/mm(2) ). Further, to test the elastography equipment on phantoms from a clinical perspective regarding the distance dependence and the influence of a heterogeneous material.
Methodological study.
Aarhus University Hospital, Denmark.
Six mid- and five full-term pregnant women.
Reference caps and phantoms with Young's moduli between 0.07 and 0.40 N/mm(2) were made of silicone and oil. By using reference caps, the approximate Young's moduli of the cervixes were calculated from strain ratios obtained by elastography.
Approximate Young's modulus of the cervix.
The recordings of the phantoms revealed that the calculation of the approximate Young's moduli became unreliable at distances above 10-15 mm from the transducer. This was further increased for a phantom which included a soft layer imitating the cervical canal. The approximate Young's modulus obtained from the anterior cervical lip was 0.08 N/mm(2) in mid-term and 0.03 N/mm(2) in full-term pregnant women (Wilcoxon rank-sum test, p = 0.01).
The reference cap constitutes a promising tool for quantitative elastography of the anterior cervical lip. Figures obtained from the posterior cervical lip are less plausible due to the distance from the transducer and the heterogeneity introduced by the cervical canal. The method has the potential to be used to supplement cervical length assessment when evaluating women at risk of preterm delivery and when planning induction of labor.
开发一种参考材料,通过计算近似组织硬度(以杨氏模量表示),实现对子宫颈的定量弹性成像。此外,从临床角度出发,对弹性成像设备在体模上进行测试,评估距离依赖性和不均匀材料的影响。
方法学研究。
丹麦奥胡斯大学医院。
六名中期妊娠和五名足月妊娠的女性。
使用硅树脂和油制作杨氏模量在 0.07 至 0.40 N/mm²之间的参考帽和体模。通过使用参考帽,从弹性成像获得的应变比计算宫颈的近似杨氏模量。
宫颈的近似杨氏模量。
体模的记录表明,当距离换能器超过 10-15 毫米时,计算宫颈的近似杨氏模量变得不可靠。对于包括模仿宫颈管的软层的体模,这种情况进一步增加。从中期妊娠到足月妊娠,从前宫颈唇获得的近似杨氏模量分别为 0.08 N/mm²和 0.03 N/mm²(Wilcoxon 秩和检验,p=0.01)。
参考帽是一种有前途的前宫颈唇定量弹性成像工具。由于距离换能器和宫颈管引入的不均匀性,从后宫颈唇获得的数值不太合理。该方法有可能在评估有早产风险的女性和计划引产时,补充宫颈长度评估。