采用 70kVp 协议的脑计算机断层血管摄影术:使用更少的造影剂和辐射剂量实现更好的血管增强效果。

Cerebral computed tomography angiography using a 70 kVp protocol: improved vascular enhancement with a reduced volume of contrast medium and radiation dose.

机构信息

Department of Radiology, Yonsei University College of Medicine, Gangnam Severance Hospital, 211 Eonjuro, Gangnam-gu, Seoul, 135-720, Korea.

出版信息

Eur Radiol. 2015 May;25(5):1421-30. doi: 10.1007/s00330-014-3540-z. Epub 2014 Dec 16.

Abstract

OBJECTIVES

To determine the feasibility of using a 70-kVp protocol compared with a 120-kVp protocol for cerebral CT angiography. An additional target was to investigate a possible reduction in the volume of contrast medium (CM) using the 70-kVp protocol.

METHODS

Attenuation value and CNR for iodine were determined at various tube voltage settings using a phantom. Sixty-nine volunteers were randomly assigned to one of three protocols: group A (120-kVp and CM 64 mL), group B (70-kVp and CM 64 mL), or group C (70-kVp and CM 40 mL). The attenuation value, SNR, and CNR of cerebral arteries, subjective image quality, and radiation dose were compared among the groups.

RESULTS

The vascular attenuation, SNR, and CNR of group B were significantly higher than those of group A. Group C had a significantly higher vascular attenuation than group A. Groups B and C were significantly better than group A with respect to subjective image quality. An effective dose of 70-kVp was 10 % lower than that of 120-kVp.

CONCLUSIONS

Using 70-kVp improved arterial enhancement, SNR, and CNR, and provided better subjective image quality, using a 10 % lower effective dose. Furthermore, the 70-kVp protocol may both reduce volume of CM by 37.5 % and improve arterial enhancement.

KEY POINTS

• Cerebral CT angiography at 70-kVp substantially improved vascular enhancement • Subjective image quality was better at 70-kVp, with lower radiation dose • The volume of contrast media can be substantially reduced at 70-kVp.

摘要

目的

确定与 120kVp 方案相比,70kVp 方案用于脑 CT 血管造影的可行性。另一个目标是研究使用 70kVp 方案可能减少造影剂(CM)体积。

方法

使用体模在不同管电压设置下测定碘的衰减值和对比噪声比(CNR)。将 69 名志愿者随机分为三组:A 组(120kVp,CM 64mL)、B 组(70kVp,CM 64mL)和 C 组(70kVp,CM 40mL)。比较三组脑动脉的衰减值、SNR 和 CNR、主观图像质量和辐射剂量。

结果

B 组的血管衰减值、SNR 和 CNR均明显高于 A 组。C 组的血管衰减值明显高于 A 组。B 组和 C 组的主观图像质量均明显优于 A 组。70kVp 的有效剂量比 120kVp 低 10%。

结论

与 120kVp 相比,使用 70kVp 可提高动脉强化、SNR 和 CNR,并提供更好的主观图像质量,同时有效剂量降低 10%。此外,70kVp 方案可将 CM 体积减少 37.5%,同时提高动脉强化。

重点

• 70kVp 脑 CT 血管造影可显著提高血管增强效果• 70kVp 时的主观图像质量更好,辐射剂量更低• 70kVp 时可大幅减少造影剂用量。