Sun Xiumei, Hu Hongmei, Zhong Zhi, Jin Yanjian, Zhang Xiaojun, Guo Yuanming
Key Laboratory of Sustainable Utilization of Technology Research for Fishery Resource of Zhejiang Province, Marine Fishery Institute of Zhejiang Province, Zhejiang Zhoushan, China.
J Sep Sci. 2015 Feb;38(4):626-33. doi: 10.1002/jssc.201400880. Epub 2015 Jan 23.
The feasibility of developing a quick, easy, efficient procedure for the simultaneous determination of organochlorinated pesticides and polychlorinated biphenyls in aquatic samples using gas chromatography with electron capture detection based on solid-phase extraction was investigated. The extraction solvent (n-hexane/acetone, cyclohexane/ethyl acetate, n-hexane/dichloromethane, n-hexane) for ultrasound-assisted solid-liquid extraction and solid-phase extraction columns (florisil, neutral alumina, acidic alumina, aminopropyl trimethoxy silane, propyl ethylenediamine, aminopropyl trimethoxy silane/propyl ethylenediamine, graphitized carbon black and silica) for cleanup procedure were optimized. The gas chromatography with electron capture detection method was validated in terms of linearity, sensitivity, reproducibility, and recovery. Mean recoveries ranged from 75 to 115% with relative standard deviations <13%. Quantification limits were 0.20-0.40 ng/g for organochlorinated pesticides and polychlorinated biphenyls. The satisfactory data demonstrated the good reproducibility of the method with relative standard deviations lower than 13%. In comparison to other related methods, this method requires less time and solvent and allows for rapid isolation of the target analytes with high selectivity. This method therefore allows for the screening of numerous samples and can also be used for routine analyses.
研究了基于固相萃取,采用带电子捕获检测的气相色谱法,开发一种快速、简便、高效的同时测定水样中有机氯农药和多氯联苯的方法的可行性。对超声辅助固液萃取的萃取溶剂(正己烷/丙酮、环己烷/乙酸乙酯、正己烷/二氯甲烷、正己烷)和净化程序所用的固相萃取柱(弗罗里硅土、中性氧化铝、酸性氧化铝、氨丙基三甲氧基硅烷、丙基乙二胺、氨丙基三甲氧基硅烷/丙基乙二胺、石墨化炭黑和硅胶)进行了优化。对带电子捕获检测的气相色谱法在线性、灵敏度、重现性和回收率方面进行了验证。平均回收率在75%至115%之间,相对标准偏差<13%。有机氯农药和多氯联苯的定量限为0.20 - 0.40 ng/g。令人满意的数据表明该方法具有良好的重现性,相对标准偏差低于13%。与其他相关方法相比,该方法所需时间和溶剂更少,能够以高选择性快速分离目标分析物。因此,该方法可用于大量样品的筛查,也可用于常规分析。