Guo Yuanming, Hu Hongmei, Li Tiejun, Xue Lijian, Zhang Xiaoning, Zhong Zhi, Zhang Yurong, Jin Yanjian
Key Laboratory of Mariculture and Enhancement of Zhejiang Province, Marine Fishery Institute of Zhejiang Province, Zhoushan, P.R. China.
Department of Mathematics, Sciences and Technology, Paine College, Augusta, GA, USA.
J Sep Sci. 2017 Aug;40(16):3279-3288. doi: 10.1002/jssc.201700315. Epub 2017 Jul 17.
A simple, rapid, and novel method has been developed and validated for determination of seven indicator polychlorinated biphenyls in water samples by gas chromatography with electron capture detection. 1 L of water samples containing 30 g of anhydrous sodium sulfate was first liquid-liquid extracted with an automated Jipad-6XB vertical oscillator using n-hexane/dichloromethane (1:1, v/v). The concentrated extract was cleaned up by dispersive solid-phase extraction with 100 mg of primary secondary amine as sorbent material. The linearity of this method ranged from 1.25 to 100 μg/L, with regression coefficients ranging between 0.9994 and 0.9999. The limits of detection were in the ng/L level, ranging between 0.2 and 0.3 ng/L. The recoveries of seven spiked polychlorinated biphenyls with external calibration method at different concentration levels in tap water, lake water, and sea water were in the ranges of 85-112, 76-116, and 72-108%, respectively, and with relative standard deviations of 3.3-4.5, 3.4-5.6, and 3.1-4.8% (n = 5), respectively. The performance of the proposed method was compared with traditional liquid-liquid extraction and solid-phase extraction clean-up methods, and comparable efficiencies were obtained. It is concluded that this method can be successfully applied for the determination of polychlorinated biphenyls in different water samples.
已开发并验证了一种简单、快速且新颖的方法,用于通过带电子捕获检测的气相色谱法测定水样中的七种指示性多氯联苯。首先,使用正己烷/二氯甲烷(1:1,v/v),通过自动Jipad - 6XB垂直振荡器对1 L含有30 g无水硫酸钠的水样进行液 - 液萃取。浓缩提取物通过以100 mg伯仲胺为吸附剂材料的分散固相萃取进行净化。该方法的线性范围为1.25至100 μg/L,回归系数在0.9994至0.9999之间。检测限在ng/L水平,范围为0.2至0.3 ng/L。采用外标法,在自来水、湖水和海水中不同浓度水平下加标的七种多氯联苯的回收率分别在85 - 112%、76 - 116%和72 - 108%范围内,相对标准偏差分别为3.3 - 4.5%、3.4 - 5.6%和3.1 - 4.8%(n = 5)。将所提出方法的性能与传统液 - 液萃取和固相萃取净化方法进行了比较,获得了相当的效率。结论是该方法可成功应用于不同水样中多氯联苯的测定。