Viñas Pilar, Campillo Natalia, Pastor-Belda Marta, Oller Ainhoa, Hernández-Córdoba Manuel
Department of Analytical Chemistry, Faculty of Chemistry, Regional Campus of International Excellence "Campus Mare Nostrum", University of Murcia, E-30100 Murcia, Spain.
Department of Analytical Chemistry, Faculty of Chemistry, Regional Campus of International Excellence "Campus Mare Nostrum", University of Murcia, E-30100 Murcia, Spain.
J Chromatogr A. 2015 Jan 9;1376:18-25. doi: 10.1016/j.chroma.2014.12.012. Epub 2014 Dec 12.
Phthalic acid esters (PEs) were preconcentrated from cleaning products, detergents and cosmetics using ultrasound assisted extraction (UAE) in the presence of acetonitrile, and then submitted to dispersive liquid-liquid microextraction (DLLME). For DLLME, 3mL of acetonitrile extract, 150μL carbon tetrachloride and 10mL aqueous solution were used. The enriched organic phase was evaporated, reconstituted with 25μL acetonitrile and injected into a liquid chromatograph with a mobile phase (acetonitrile:10mM ammonium acetate, pH 4) under gradient elution. Detection was carried out using both diode-array (DAD) and electrospray-ion trap-tandem mass spectrometry (ESI-IT-MS/MS) in the multiple reaction monitoring mode (MRM) of the positive fragment ions. Quantification was carried out using matrix-matched standards. Detection limits were in the range 0.04-0.45ngmL(-1) for the six PEs considered. The recoveries obtained were in the 84-124% range, with RSDs lower than 10%. Thirty three different cleaning products were analyzed. The most frequently found compound was diethyl phthalate.
在乙腈存在的情况下,采用超声辅助萃取(UAE)从清洁产品、洗涤剂和化妆品中预富集邻苯二甲酸酯(PEs),然后进行分散液液微萃取(DLLME)。对于DLLME,使用3mL乙腈萃取液、150μL四氯化碳和10mL水溶液。将富集的有机相蒸发,用25μL乙腈复溶,然后注入配备流动相(乙腈:10mM醋酸铵,pH 4)的液相色谱仪中进行梯度洗脱。采用二极管阵列(DAD)和电喷雾离子阱串联质谱(ESI-IT-MS/MS)在正离子碎片的多反应监测模式(MRM)下进行检测。使用基质匹配标准品进行定量。所考虑的六种PEs的检测限在0.04 - 0.45ngmL⁻¹范围内。获得的回收率在84 - 124%范围内,相对标准偏差低于10%。分析了33种不同的清洁产品。最常检测到的化合物是邻苯二甲酸二乙酯。