采用MAVRIC SL在3-T磁共振成像中减少髋关节置换术患者的金属伪影

Metal artifact reduction with MAVRIC SL at 3-T MRI in patients with hip arthroplasty.

作者信息

Choi Soo-Jung, Koch Kevin M, Hargreaves Brian A, Stevens Kathryn J, Gold Garry E

机构信息

1 Department of Radiology, Asan Foundation, Gangneung Asan Hospital, University of Ulsan, College of Medicine, Gangneung-si, Gangwon-do, Korea.

出版信息

AJR Am J Roentgenol. 2015 Jan;204(1):140-7. doi: 10.2214/AJR.13.11785.

Abstract

OBJECTIVE

The objective of our study was to compare the multiacquisition variable-resonance image combination selective (MAVRIC SL) sequence with the 2D fast spin-echo (FSE) sequence for metal artifact reduction on 3-T MRI in patients with hip arthroplasty (HA).

MATERIALS AND METHODS

Matched 2D FSE and MAVRIC SL images of 21 hips (19 patients with HA) were included in the study group. Paired image sets, composed of 13 coronal and 12 axial slices (total, 25 image sets), of the 21 hips were evaluated. For quantitative analysis, the artifact area was measured at the level of the hip and femur. For qualitative analysis, two musculoskeletal radiologists independently compared paired 2D FSE and MAVRIC SL sets in terms of artifacts, depiction of anatomic detail, level of diagnostic confidence, and detection of abnormal findings.

RESULTS

The measured artifact area was significantly smaller (p < 0.05) on MAVRIC SL than 2D FSE at both the level of hip (59.9% reduction with MAVRIC SL) and femur (31.3% reduction with MAVRIC SL). The artifact score was also significantly decreased (p < 0.0001) with MAVRIC SL compared with 2D FSE for both reviewers. The hip joint capsule and the muscle and tendon attachment sites of the obturator externus and iliopsoas muscles were better depicted with MAVRIC SL than 2D FSE (p < 0.0125). Abnormal findings were significantly better shown on MAVRIC SL imaging compared with 2D FSE imaging (p < 0.0001).

CONCLUSION

The MAVRIC SL sequence can significantly reduce metal artifact on 3-T MRI compared with the 2D FSE sequence and can increase diagnostic confidence of 3-T MRI in patients with total HA.

摘要

目的

本研究的目的是比较多采集可变共振图像组合选择性(MAVRIC SL)序列与二维快速自旋回波(FSE)序列在3-T磁共振成像(MRI)上减少髋关节置换术(HA)患者金属伪影的效果。

材料与方法

研究组纳入了21个髋关节(19例HA患者)的匹配二维FSE和MAVRIC SL图像。对这21个髋关节的成对图像集进行评估,每个图像集由13个冠状位和12个轴位切片组成(共25个图像集)。进行定量分析时,在髋关节和股骨水平测量伪影面积。进行定性分析时,两名肌肉骨骼放射科医生独立比较成对的二维FSE和MAVRIC SL图像集,比较内容包括伪影、解剖细节显示、诊断置信度水平以及异常发现的检测。

结果

在髋关节水平(MAVRIC SL减少59.9%)和股骨水平(MAVRIC SL减少31.3%),MAVRIC SL测量的伪影面积均显著小于二维FSE(p < 0.05)。与二维FSE相比,两名评估者对MAVRIC SL的伪影评分也显著降低(p < 0.0001)。MAVRIC SL比二维FSE能更好地显示髋关节囊以及闭孔外肌和髂腰肌的肌肉和肌腱附着部位(p < 0.0125)。与二维FSE成像相比,MAVRIC SL成像能更显著地显示异常发现(p < 0.0001)。

结论

与二维FSE序列相比,MAVRIC SL序列可显著减少3-T MRI上的金属伪影,并可提高3-T MRI对全髋关节置换术患者的诊断置信度。

相似文献

引用本文的文献

9
Imaging approach to prosthetic joint infection.人工关节感染的影像学方法。
Skeletal Radiol. 2024 Oct;53(10):2023-2037. doi: 10.1007/s00256-023-04546-7. Epub 2023 Dec 22.

本文引用的文献

5
Magnetic resonance imaging of the postoperative hip.术后髋关节的磁共振成像。
J Magn Reson Imaging. 2012 May;35(5):1013-25. doi: 10.1002/jmri.23523.
8
Metal-induced artifacts in MRI.磁共振成像中的金属伪影。
AJR Am J Roentgenol. 2011 Sep;197(3):547-55. doi: 10.2214/AJR.11.7364.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验