Metal artefact suppression at 3 T MRI: comparison of MAVRIC-SL with conventional fast spin echo sequences in patients with Hip joint arthroplasty.

作者信息

Kretzschmar Martin, Nardo Lorenzo, Han Misung M, Heilmeier Ursula, Sam Craig, Joseph Gabby B, Koch Kevin M, Krug Roland, Link Thomas M

机构信息

Musculoskeletal Quantitative Imaging Research Group, Department of Radiology & Biomedical Imaging, University of California San Francisco, 185 Berry Street, San Francisco, CA, 94107, USA,

出版信息

Eur Radiol. 2015 Aug;25(8):2403-11. doi: 10.1007/s00330-015-3628-0. Epub 2015 Feb 14.

Abstract

OBJECTIVE

The aim of our study was to evaluate the clinical feasibility and diagnostic value of a new MRI metal artefact reduction pulse sequence called MAVRIC-SL in a 3 T MRI environment.

SUBJECTS AND METHODS

Two MAVRIC-SL sequences obtained in 61 patients with symptomatic total hip replacement were compared with standard FSE-STIR sequences optimized for imaging around metal. Artefact size was measured on the slice of greatest extent. Image quality, fat saturation, image distortion, visibility of anatomical structures, and detectability of joint abnormalities were visually assessed and graded on qualitative scales. Differences between MAVRIC-SL and FSE sequences were tested with the Wilcoxon signed-rank test.

RESULTS

MAVRIC-SL sequences at 3 T showed significantly smaller metal artefacts compared to FSE-STIR sequences (p < 0.0001). The general image quality of MAVRIC-SL sequences was reduced with regard to spatial resolution, noise and contrast (p = 0.001), and fat saturation (p < 0.0001). The reduction of artefact size and image distortion significantly improved visualization of joint anatomy (p < 0.0001) and diagnostic confidence regarding implant-associated abnormalities (p = 0.0075 to <0.0001).

CONCLUSION

Although the image quality of MAVRIC-SL sequences is limited at 3 T, its clinical application is feasible and provides important additional diagnostic information for the workup of patients with symptomatic hip replacement through substantially reduced metal artefacts.

KEY POINTS

• Metal artefacts compromise imaging of total hip replacement with MRI. • Metal artefacts are aggravated with 3 Tesla MRI. • MAVRIC-SL is a technique to suppress metal artefacts. • MAVRIC-SL effectively reduces metal artefacts at 3 Tesla and improves diagnostic quality.

摘要

目的

本研究旨在评估一种名为MAVRIC-SL的新型磁共振成像(MRI)金属伪影减少脉冲序列在3T MRI环境中的临床可行性和诊断价值。

受试者与方法

对61例有症状的全髋关节置换患者采用两种MAVRIC-SL序列,并与针对金属周围成像优化的标准快速自旋回波短TI反转恢复(FSE-STIR)序列进行比较。在伪影最大范围的层面上测量伪影大小。对图像质量、脂肪抑制、图像变形、解剖结构可见性以及关节异常的可检测性进行视觉评估,并根据定性量表进行分级。采用Wilcoxon符号秩检验来检验MAVRIC-SL序列与FSE序列之间的差异。

结果

与FSE-STIR序列相比,3T时的MAVRIC-SL序列显示出明显更小的金属伪影(p < 0.0001)。MAVRIC-SL序列的总体图像质量在空间分辨率、噪声和对比度方面有所降低(p = 0.001),脂肪抑制方面也有所降低(p < 0.0001)。伪影大小和图像变形的减少显著改善了关节解剖结构的可视化(p < 0.0001)以及对植入物相关异常的诊断信心(p = 0.0075至<0.0001)。

结论

尽管MAVRIC-SL序列在3T时图像质量有限,但其临床应用是可行的,并且通过大幅减少金属伪影,为有症状髋关节置换患者的检查提供了重要的额外诊断信息。

关键点

• 金属伪影会影响MRI对全髋关节置换的成像。• 3特斯拉MRI会加重金属伪影。• MAVRIC-SL是一种抑制金属伪影的技术。• MAVRIC-SL在3特斯拉时能有效减少金属伪影并提高诊断质量。