We found previously that albumin-calibrated total protein in certified reference materials causes unacceptable positive bias in analysis of human sera. The simplest way to cure this defect is the use of human-based serum/plasma standards calibrated by the Kjeldahl method. Such standards, commutative with serum samples, will compensate for bias caused by lipids and bilirubin in most human sera. To find a suitable primary reference procedure for total protein in reference materials, we reviewed Kjeldahl methods adopted by laboratory medicine. We found two methods recommended for total protein in human samples: an indirect analysis based on total Kjeldahl nitrogen corrected for its nonprotein nitrogen and a direct analysis made on isolated protein precipitates. The methods found will be assessed in a subsequent article.
我们之前发现,认证参考物质中的白蛋白校准总蛋白会导致分析人血清时出现不可接受的正偏倚。纠正这一缺陷的最简单方法是使用经凯氏定氮法校准的基于人源的血清/血浆标准品。这些与血清样本兼容的标准品将补偿大多数人血清中脂质和胆红素引起的偏差。为了找到参考材料中总蛋白的合适的基准参考程序,我们回顾了实验室医学中采用的凯氏定氮法。我们发现了两种推荐用于人样本总蛋白的方法:一种基于总凯氏氮并校正其非蛋白氮的间接分析,以及一种对分离的蛋白质沉淀物进行的直接分析。后续的文章中将对这些方法进行评估。