Lin Robert, Skandarajah Arunan, Gerver Rachel E, Neira Hector D, Fletcher Daniel A, Herr Amy E
Department of Bioengineering, UC Berkeley, Berkeley, CA 94720 USA.
Lab Chip. 2015 Mar 21;15(6):1488-96. doi: 10.1039/c4lc01370k.
Immunochromatographic assays are a cornerstone tool in disease screening. To complement existing lateral flow assays (based on wicking flow) we introduce a lateral flow format that employs directed electrophoretic transport. The format is termed a "lateral e-flow assay" and is designed to support multiplexed detection using immobilized reaction volumes of capture antigen. To fabricate the lateral e-flow device, we employ mask-based UV photopatterning to selectively immobilize unmodified capture antigen along the microchannel in a barcode-like pattern. The channel-filling polyacrylamide hydrogel incorporates a photoactive moiety (benzophenone) to immobilize capture antigen to the hydrogel without a priori antigen modification. We report a heterogeneous sandwich assay using low-power electrophoresis to drive biospecimen through the capture antigen barcode. Fluorescence barcode readout is collected via a low-resource appropriate imaging system (CellScope). We characterize lateral e-flow assay performance and demonstrate a serum assay for antibodies to the hepatitis C virus (HCV). In a pilot study, the lateral e-flow assay positively identifies HCV+ human sera in 60 min. The lateral e-flow assay provides a flexible format for conducting multiplexed immunoassays relevant to confirmatory diagnosis in near-patient settings.
免疫层析测定法是疾病筛查中的一项基础工具。为补充现有的侧向流动测定法(基于毛细流动),我们引入了一种采用定向电泳传输的侧向流动形式。这种形式被称为“侧向电子流测定法”,旨在支持使用固定化捕获抗原反应体积进行多重检测。为制造侧向电子流装置,我们采用基于掩膜的紫外光图案化技术,以条形码状图案沿着微通道选择性地固定未修饰的捕获抗原。通道填充的聚丙烯酰胺水凝胶包含一个光活性部分(二苯甲酮),无需事先对抗原进行修饰即可将捕获抗原固定到水凝胶上。我们报告了一种使用低功率电泳驱动生物样本通过捕获抗原条形码的异质夹心测定法。通过低资源适用的成像系统(CellScope)收集荧光条形码读数。我们对侧向电子流测定法的性能进行了表征,并展示了一种针对丙型肝炎病毒(HCV)抗体的血清测定法。在一项初步研究中,侧向电子流测定法在60分钟内阳性鉴定出HCV阳性的人血清。侧向电子流测定法为在床旁进行与确诊相关的多重免疫测定提供了一种灵活的形式。