Suppr超能文献

高速双气动玻璃体切割探头相关吸引能力的比较。

Comparison of attraction capabilities associated with high-speed, dual-pneumatic vitrectomy probes.

作者信息

Dugel Pravin U, Abulon Dina J K, Dimalanta Ramon

机构信息

*Retinal Consultants of AZ, Ltd, Phoenix, Arizona; and †Global Medical Affairs, Alcon Research, Ltd., Irvine, California.

出版信息

Retina. 2015 May;35(5):915-20. doi: 10.1097/IAE.0000000000000411.

Abstract

PURPOSE

To measure membrane attraction capabilities of enhanced 27-gauge, enhanced 25-gauge, and 23-gauge vitrectomy probes under various parameters.

METHODS

A membrane-on-cantilever apparatus was used to measure membrane attraction for enhanced 27-, enhanced 25-, and 23-gauge UltraVit probes (n = 6 for each). The following parameters were evaluated: effects of cut rates and duty cycles on membrane attraction distances, and flow rates and vacuum levels required to attract a membrane at a fixed distance.

RESULTS

The enhanced 27-gauge probe had the shortest attraction distance across all cutting speeds and duty cycles. To attract a membrane at a fixed distance, increasing vacuum was necessary with higher cutting rates and smaller probe gauges but flow rate remained relatively constant. The biased open duty cycle was associated with a longer attraction distance than 50/50 or biased closed modes.

CONCLUSION

The shorter membrane attraction distance of the enhanced 27-gauge probe versus 23-gauge and enhanced 25-gauge probes may permit greater membrane dissection precision while providing improved access to small tissue planes. Equivalent fluid flow capabilities of the 27-gauge probe compared with the 23-gauge and 25-gauge probes may provide efficient aspiration. Surgeon selection of duty cycle modes may improve intraoperative fluid control and expand the cutter utility as a multifunctional tool.

摘要

目的

测量在各种参数下增强型27G、增强型25G和23G玻璃体切割探头的膜吸引能力。

方法

使用悬臂式膜装置测量增强型27G、增强型25G和23G UltraVit探头的膜吸引力(每种探头n = 6)。评估以下参数:切割速率和占空比对膜吸引距离的影响,以及在固定距离吸引膜所需的流速和真空度。

结果

在所有切割速度和占空比下,增强型27G探头的吸引距离最短。为了在固定距离吸引膜,随着切割速率提高和探头规格变小,需要增加真空度,但流速保持相对恒定。与50/50或偏置关闭模式相比,偏置打开占空比与更长的吸引距离相关。

结论

与23G和增强型25G探头相比,增强型27G探头的膜吸引距离更短,这可能在提供更好的小组织平面进入条件的同时,实现更高的膜切割精度。27G探头与23G和25G探头相当的流体流动能力可能提供有效的抽吸。外科医生选择占空比模式可能改善术中液体控制,并扩大切割器作为多功能工具的用途。