A new facility for the synchrotron radiation-based calibration of transfer radiation sources in the ultraviolet and vacuum ultraviolet spectral range.

作者信息

Thornagel Reiner, Fliegauf Rolf, Klein Roman, Kroth Simone, Paustian Wolfgang, Richter Mathias

机构信息

Physikalisch-Technische Bundesanstalt (PTB), Berlin, Germany.

出版信息

Rev Sci Instrum. 2015 Jan;86(1):013106. doi: 10.1063/1.4905187.

Abstract

The Physikalisch-Technische Bundesanstalt (PTB) has a long tradition in the calibration of radiation sources in the ultraviolet and vacuum ultraviolet spectral range, with traceability to calculable synchrotron radiation. Within this context, new instrumentation in the PTB laboratory at the Metrology Light Source (MLS) has been put into operation that opens up extended and improved calibration possibilities. A new facility for radiation source calibrations has been set up in the spectral range from 7 nm to 400 nm based on a combined normal incidence-grazing incidence monochromator. The facility can be used for the calibration of transfer sources in terms of spectral radiant intensity or mean spectral radiance, with traceability to the MLS primary source standard. We describe the design and performance of the experimental station and give examples of some commissioning results.