Messinger Maren, Albert Frank, Rohde Nils
Clin Lab. 2014;60(12):2007-13. doi: 10.7754/clin.lab.2014.140403.
High sensitive troponin assay methods used in emergency departments and those used in hospital laboratories are needed with low imprecision of the analytical method aligned along the cut-off.
Presented is a method for adjusting the Troponin I (TnI) assay on the AQT90 FLEX analyzer (Radiometer) to the UniCel DxI 800 immunoassay system (Beckman Coulter) by correlation analysis using the Passing-Bablok method, followed by verification using receiver operating characteristic (ROC) analysis. This allows a uniform TnI cut-off to be applied to the use of different immunoassay systems in both the emergency department and in hospital laboratory settings. Finally, an imprecision profile of the TnI assay methods is applied to present a tool that allows for the discrimination of significant TnI results from imprecise measurements produced by the immunoassay system during repeated measurements.
For TnI, the equation of the best-fit lines with which all AQT90 FLEX results can be adapted to the UniCel DxL 800 is: UniCel DxI = 4.9524 AQT90 FLEX + 0.003. The imprecision of TnI can be calculated using the mathematical relationship described by Reed et al. from the assay's coefficient of variation and the difference between the first TnI result to the second TnI result.
It is recommended to use a slope of 4.9524 and an intercept of 0.003 to adjust the TnI values of the AQT90 FLEX to the UniCel DxI 800 assay systems. Clinically relevant, i.e., significant TnI changes (rise or fall), can be reliably interpreted by accounting for the disparate results and the imprecision profile of the TnI assay method by applying the mathematical relationship described by Reed.
急诊科和医院实验室使用的高敏肌钙蛋白检测方法需要在临界值附近具有低分析不精密度。
本文介绍了一种通过使用Passing-Bablok方法进行相关分析,将AQT90 FLEX分析仪(Radiometer公司)上的肌钙蛋白I(TnI)检测方法调整至UniCel DxI 800免疫分析系统(Beckman Coulter公司)的方法,随后使用受试者工作特征(ROC)分析进行验证。这使得在急诊科和医院实验室环境中,能够将统一的TnI临界值应用于不同免疫分析系统的使用。最后,应用TnI检测方法的不精密度图谱,以呈现一种工具,该工具能够区分免疫分析系统在重复测量过程中产生的不精确测量结果与显著的TnI结果。
对于TnI,可使所有AQT90 FLEX结果适配至UniCel DxL 800的最佳拟合线方程为:UniCel DxI = 4.9524×AQT90 FLEX + 0.003。TnI的不精密度可使用Reed等人描述的数学关系,根据检测的变异系数以及第一个TnI结果与第二个TnI结果之间的差异来计算。
建议使用斜率4.9524和截距0.003将AQT90 FLEX的TnI值调整至UniCel DxI 800检测系统。通过应用Reed描述的数学关系,考虑到不同的结果以及TnI检测方法的不精密度图谱,可对临床相关的即显著的TnI变化(升高或降低)进行可靠解读。