Bobály Balázs, Beck Alain, Fekete Jenő, Guillarme Davy, Fekete Szabolcs
Department of Inorganic and Analytical Chemistry, Budapest University of Technology and Economics, Szt. Gellért tér 4, 1111 Budapest, Hungary.
Center of Immunology Pierre Fabre, 5 Avenue Napoléon III, BP 60497, 74160 Saint-Julien-en-Genevois, France.
Talanta. 2015 May;136:60-7. doi: 10.1016/j.talanta.2014.12.006. Epub 2015 Jan 13.
Trifluoroacetic acid (TFA) is commonly used as mobile phase additive for the analysis of proteins in reversed phase liquid chromatography (RPLC). Due to its interesting features, it provides symmetrical and narrow peak shapes for proteins, but decreases mass spectrometric sensitivity through ion-pairing and spray destabilizing. Since RPLC-MS is an important technique for the characterization of proteins, some alternative MS-compatible mobile phases may be required. The aim of this study was to evaluate various acidic and basic mobile phase additives for the LC-MS analysis of therapeutic proteins possessing molecular weight between 5 and 150kDa. At the end, 10mM formate buffer pH 3 was found to be the most promising alternative, since it provided acceptable peak shapes in most cases, together with an average improvement of MS sensitivity by 5-times, compared to TFA.
三氟乙酸(TFA)常用于反相液相色谱(RPLC)中蛋白质分析的流动相添加剂。由于其有趣的特性,它为蛋白质提供对称且窄的峰形,但通过离子对作用和喷雾不稳定作用降低了质谱灵敏度。由于RPLC-MS是蛋白质表征的重要技术,可能需要一些替代的与质谱兼容的流动相。本研究的目的是评估各种酸性和碱性流动相添加剂用于分子量在5至150 kDa之间的治疗性蛋白质的LC-MS分析。最后,发现10 mM pH 3的甲酸盐缓冲液是最有前景的替代品,因为在大多数情况下它提供了可接受的峰形,与TFA相比,质谱灵敏度平均提高了5倍。