Abstract

PURPOSE

To evaluate the visual acuity agreement between a standard back-illuminated Early Treatment Diabetic Retinopathy Study (ETDRS) chart and a handheld internally illuminated ETDRS chart.

METHODS

Two-center prospective study. Seventy patients (134 eyes) with retinal pathology were enrolled between October 2012 and August 2013. Visual acuity was measured using both the ETDRS chart and the handheld device by masked independent examiners after best protocol refraction. Examination was performed in the same room under identical illumination and testing conditions.

RESULTS

The mean number of letters seen was 63.0 (standard deviation: 19.8 letters) and 61.2 letters (standard deviation: 19.1 letters) for the ETDRS chart and handheld device, respectively. Mean difference per eye between the ETDRS and handheld device was 1.8 letters. A correlation coefficient (r) of 0.95 demonstrated a positive linear correlation between ETDRS chart and handheld device measured acuities. Intraclass correlation coefficient was performed to assess the reproducibility of the measurements made by different observers measuring the same quantity and was calculated to be 0.95 (95% confidence interval: 0.93-0.96). Agreement was independent of retinal disease.

CONCLUSION

The strong correlation between measured visual acuity using the ETDRS and handheld equivalent suggests that they may be used interchangeably, with accurate measurements. Potential benefits of this device include convenience and portability, as well as the ability to assess ETDRS visual acuity without a dedicated testing lane.

摘要

目的

评估标准背照式早期糖尿病视网膜病变研究(ETDRS)视力表与手持式内部照明ETDRS视力表之间的视力一致性。

方法

两中心前瞻性研究。2012年10月至2013年8月期间纳入了70例(134只眼)患有视网膜病变的患者。在最佳验光方案后,由蒙面独立检查者使用ETDRS视力表和手持式设备测量视力。检查在同一房间内相同照明和测试条件下进行。

结果

ETDRS视力表和手持式设备每只眼所见字母的平均数分别为63.0个(标准差:19.8个字母)和61.2个字母(标准差:19.1个字母)。ETDRS视力表与手持式设备每只眼的平均差异为1.8个字母。相关系数(r)为0.95,表明ETDRS视力表与手持式设备测量的视力之间呈正线性相关。进行组内相关系数分析以评估不同观察者测量同一量值的测量结果的可重复性,计算得出为0.95(95%置信区间:0.93 - 0.96)。一致性与视网膜疾病无关。

结论

使用ETDRS视力表和手持式等效设备测量的视力之间的强相关性表明,它们在测量准确的情况下可以互换使用。该设备的潜在优点包括便利性和便携性,以及无需专用测试通道即可评估ETDRS视力的能力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验