Rofouei Mohammad K, Amiri Nayereh, Ghasemi Jahan B
Faculty of Chemistry, Kharazmi University, Tehran, Iran.
Environ Monit Assess. 2015 Mar;187(3):149. doi: 10.1007/s10661-015-4378-9. Epub 2015 Feb 27.
A simple method was developed for the determination of Au(III) and Pt(IV) contents in aqueous samples after preconcentration. The method was based on the sorption of analytes as 2-amino-5-mercapto-1,3,4-thiadiazol complexes onto graphene oxide and subsequent direct determination by wavelength dispersive X-ray fluorescence (WDXRF). The optimization step was carried out using two-level full-factorial and Box-Behnken designs. The effects of four variables (pH, ligand mass, sonication time, and temperature) were studied by a full-factorial design to find significant variables and their interactions. Results of two-level full-factorial design for Au extraction showed that the factors: pH, ligand mass, temperature of sonication beside the interaction of pH-ligand mass, and interaction sonication temperature-ligand mass were significant. For Pt, the results revealed pH, ligand mass, sonication time, and interaction of pH-ligand mass were statistically significant. Box-Behnken matrix design was applied to determine the optimum level of significant parameters for extraction of two analytes simultaneously. The optimum values of the factors were pH 2.5, 0.9 mL ligand solution, 56 min sonication time and 15 °C temperature. The limits of detection (LOD) were found to be 8 ng mL(-1) for Au and 6 ng mL(-1) for Pt. The adsorption capacity for Au and Pt were 115 and 169 μg mg(-1), respectively. The relative standard deviation (RSD) was lower than 1.4 % (n = 5), and the extraction percentage was more than 95 % for both elements. The method was validated by determination of Au and Pt in spiked water samples and certified reference standard materials.
开发了一种简单的方法,用于在预富集后测定水性样品中的金(III)和铂(IV)含量。该方法基于将分析物作为2-氨基-5-巯基-1,3,4-噻二唑配合物吸附到氧化石墨烯上,随后通过波长色散X射线荧光光谱法(WDXRF)进行直接测定。使用二级全因子设计和Box-Behnken设计进行优化步骤。通过全因子设计研究了四个变量(pH值、配体质量、超声处理时间和温度)的影响,以找出显著变量及其相互作用。金萃取的二级全因子设计结果表明,因素:pH值、配体质量、超声处理温度以及pH值-配体质量的相互作用和超声处理温度-配体质量的相互作用具有显著性。对于铂,结果表明pH值、配体质量、超声处理时间以及pH值-配体质量的相互作用具有统计学显著性。应用Box-Behnken矩阵设计来确定同时萃取两种分析物的显著参数的最佳水平。因素的最佳值为pH 2.5、0.9 mL配体溶液、56分钟超声处理时间和15°C温度。发现金的检测限(LOD)为8 ng mL⁻¹,铂的检测限为6 ng mL⁻¹。金和铂的吸附容量分别为115和169 μg mg⁻¹。相对标准偏差(RSD)低于1.4%(n = 5),两种元素的萃取率均超过95%。通过测定加标水样和有证参考标准物质中的金和铂对该方法进行了验证。