Mahara Aditya, Khan Shadab, Murphy Ethan K, Schned Alan R, Hyams Elias S, Halter Ryan J
IEEE Trans Med Imaging. 2015 Jul;34(7):1590-1601. doi: 10.1109/TMI.2015.2407833. Epub 2015 Feb 27.
Radially configured microendoscopic electrical impedance probes intended for intraoperative surgical margin assessment during robot-assisted laparoscopic prostatectomy (RALP) were examined through simulation, bench-top experimentation, and ex vivo tissue studies. Three probe designs with 8, 9, and 17 electrodes, respectively, were analyzed through finite element method based simulations. One mm diameter spherical inclusions ( σinclusion = 1 S/m) are positioned at various locations within a hemispherical background ( σbackground = 0.1 S/m) of radius 5 mm. An 8-electrode configuration is not able to localize the inclusion at these positions while 9 and 17-electrode configurations are able to accurately reconstruct the inclusion at maximum depth of 1 mm and 3 mm, respectively. All three probe designs were constructed and evaluated using saline phantoms and ex vivo porcine and human prostate tissues. The 17-electrode probe performed best in saline phantom studies, accurately reconstructing high contrast, 1-mm-diameter metal cylindrical inclusions in a saline bath ( σsaline = 0.1 S/m) with a position and area error of 0.46 mm and 0.84 mm2, respectively. Additionally, the 17-electrode probe was able to adequately distinguish cancerous from benign tissues in three ex vivo human prostates. Simulations, bench-top saline experiments, and ex vivo tissue sampling suggest that for intraoperative surgical margin assessment during RALP, the 17-electrode probe (as compared to an 8 and 9 electrode probe) will be necessary to provide sufficient accuracy and sensitivity.
通过模拟、台式实验和离体组织研究,对用于机器人辅助腹腔镜前列腺切除术(RALP)术中手术切缘评估的径向配置微内镜电阻抗探头进行了检查。分别通过基于有限元方法的模拟分析了三种探头设计,电极数量分别为8个、9个和17个。直径1mm的球形内含物(σ内含物 = 1 S/m)被放置在半径5mm的半球形背景(σ背景 = 0.1 S/m)内的不同位置。8电极配置无法在这些位置定位内含物,而9电极和17电极配置能够分别在最大深度1mm和3mm处准确重建内含物。所有三种探头设计均使用盐水体模以及离体猪和人类前列腺组织进行构建和评估。17电极探头在盐水体模研究中表现最佳,能够在盐水浴(σ盐水 = 0.1 S/m)中准确重建高对比度、直径1mm的金属圆柱形内含物,位置误差和面积误差分别为0.46mm和0.84mm²。此外,17电极探头能够在三个离体人类前列腺中充分区分癌组织和良性组织。模拟、台式盐水实验和离体组织采样表明,对于RALP术中手术切缘评估,17电极探头(与8电极和9电极探头相比)对于提供足够的准确性和灵敏度是必要的。