Umemura Shin-ichiro
Graduate School of Biomedical Engineering, Tohoku University , Sendai , Japan.
Int J Hyperthermia. 2015 Mar;31(2):216-21. doi: 10.3109/02656736.2015.1008059. Epub 2015 Mar 10.
Air-backed transducers have been employed for thermal ultrasonic treatment including both ablation and hyperthermia because the power efficiency rather than the bandwidth is a main concern, unlike a typical imaging transducer working in a pulse mode. The characteristic of an air-backed piezoelectric transducer with a matching layer is analysed, and the role and choice of the matching layer is discussed. An element size of a focused array transducer, appropriate for such thermal treatment, is then estimated, and the characteristic of a piezoceramic transducer element of such a size was numerically analysed using a finite element code. The characteristic of a piezocomposite transducer element is also numerically analysed and its suitability to such a therapeutic array transducer is discussed.