Large Thomas M, Agel Julie, Holtzman Daniel J, Benirschke Stephen K, Krieg James C
*Orthopaedic Trauma Services, Mission Hospital, Asheville, NC; †Department of Orthopaedic Surgery and Sports Medicine, Harborview Medical Center, Seattle, WA; and ‡Rothman Institute at Thomas Jefferson University, Philadelphia, PA.
J Orthop Trauma. 2015 Jul;29(7):316-21. doi: 10.1097/BOT.0000000000000317.
To determine whether interobserver technical variations and errors in the measurement of compartment pressures may affect measurement accuracy.
Four above-knee cadaveric specimens were used to create a consistent model of lower leg compartment syndrome. Thirty-eight physicians examined the limbs and measured 4 compartment pressures using the Intra-Compartmental Pressure Monitor (Stryker Orthopaedics). They were observed for correct assembly and use of the monitor. Measurements obtained were compared with known pressures.
Of the total number of compartment measurements, 31% were made using the correct technique, 39% were made with lesser errors in technique, and 30% were made with catastrophic errors. Only 60% of measurements made with the correct technique were within 5 mm Hg of the standard pressure. Accuracy dropped to 42% for measurements taken with small errors in technique and 22% when a catastrophic error was committed.
Variations in use of a commercially available pressure monitor exist, and errors are common. Proper use improved accuracy, but even with proper technique, 40% of the measurements were >5 mm Hg from the actual pressure. Based on our data, measurement accuracy with this device should be questioned and viewed within a range. Regular review and education of technique is strongly recommended.
确定观察者之间的技术差异以及间室压力测量中的误差是否会影响测量准确性。
使用四个膝上尸体标本建立一致的小腿骨筋膜室综合征模型。38名医生检查这些肢体,并使用骨筋膜室内压力监测仪(史赛克骨科公司)测量4个间室的压力。观察他们对监测仪的正确组装和使用情况。将获得的测量值与已知压力进行比较。
在所有间室测量中,31%采用了正确技术,39%技术误差较小,30%存在严重误差。采用正确技术进行的测量中,只有60%在标准压力的5毫米汞柱范围内。技术误差较小时测量的准确性降至42%,出现严重误差时则降至22%。
市售压力监测仪的使用存在差异,误差很常见。正确使用可提高准确性,但即使技术正确,仍有40%的测量值与实际压力相差>5毫米汞柱。根据我们的数据,该设备的测量准确性值得质疑,应在一定范围内看待。强烈建议定期对技术进行审查和培训。