Fania Claudio, Benetti Elisabetta, Palatini Paolo
Department of Medicine, University of Padova, Padua, Italy.
Blood Press Monit. 2015 Aug;20(4):237-40. doi: 10.1097/MBP.0000000000000117.
The aim of this study was to determine the accuracy of the A&D BP UB-543 wrist device for home blood pressure (BP) measurement according to the International Protocol of the European Society of Hypertension. The A&D BP UB-543 monitor is provided with a correct position guidance (CPG) indicator that ensures the correct placement of the wrist at the heart level. Device evaluation was carried out in 33 patients with the CPG indicator on. The mean age of the patients was 53.1±16.4 years, the mean systolic BP was 141.8±25.1 mmHg (range 84 : 196), the mean diastolic BP was 88.2±14.5 mmHg (range 34 : 132), the mean arm circumference was 29.0±3.6 cm (range 21 : 38), and the mean wrist circumference was 17.5±1.4 cm (range 15 : 20). The protocol requirements were followed precisely. The device passed all requirements, fulfilling the standards of the protocol. On average, the device overestimated the systolic BP by 1.1±2.9 mmHg and underestimated diastolic BP by 0.1±3.0 mmHg. These data show that the A&D BP UB-543 wrist device used correctly with the CPG indicator on met the requirements for validation by the International Protocol and can be recommended for clinical use in the adult population.
本研究旨在根据欧洲高血压学会的国际协议,确定A&D BP UB - 543腕式设备用于家庭血压测量的准确性。A&D BP UB - 543监测仪配备了正确位置引导(CPG)指示器,可确保将手腕正确放置于心脏水平。在33名开启CPG指示器的患者中进行了设备评估。患者的平均年龄为53.1±16.4岁,平均收缩压为141.8±25.1 mmHg(范围84∶196),平均舒张压为88.2±14.5 mmHg(范围34∶132),平均臂围为29.0±3.6 cm(范围21∶38),平均腕围为17.5±1.4 cm(范围15∶20)。严格遵循了协议要求。该设备通过了所有要求,符合协议标准。平均而言,该设备收缩压高估1.1±2.9 mmHg,舒张压低估0.1±3.0 mmHg。这些数据表明,开启CPG指示器正确使用的A&D BP UB - 543腕式设备符合国际协议的验证要求,可推荐用于成年人群的临床使用。