Muhogora W, Ngoye W, Byorushengo E, Lwakatare F, Kalambo C
Tanzania Atomic Energy Commission, Block J, Njiro, Arusha, Tanzania
Tanzania Atomic Energy Commission, Block J, Njiro, Arusha, Tanzania.
Radiat Prot Dosimetry. 2016 Feb;168(2):253-60. doi: 10.1093/rpd/ncv021. Epub 2015 Mar 19.
The aim of this study was to determine the radiation doses to paediatric patients of different age groups at three large hospitals for optimisation purposes. The entrance surface air kerma (ESAK) values were determined from the measured X-ray output values using calibrated ionisation chamber, TW 233612 and clinical patient parameters. The air kerma-area product (KAP) values were measured using a calibrated Diamentor E2 system. The volume computed tomography dose index (CTDIvol) and dose length product (DLP) values were obtained from the computed tomography (CT) equipment verified by a calibrated CT chamber, Unifors Xi CT. Irrespective of age groups, the results show that the median ESAK values ranged from 62.6 to 248.1 µGy. The median KAP values ranged from 135.6 to 1612 µGy cm(2), while the median DLP values ranged from 119.1 to 600 mGy cm. Analysis of the results indicates that optimisation can be achieved through good practice awareness and patient dose and image quality evaluations.
本研究的目的是确定三家大型医院不同年龄组儿科患者的辐射剂量,以进行优化。使用校准电离室、TW 233612和临床患者参数,根据测量的X射线输出值确定入射表面空气比释动能(ESAK)值。使用校准的Diamentor E2系统测量空气比释动能面积乘积(KAP)值。容积计算机断层扫描剂量指数(CTDIvol)和剂量长度乘积(DLP)值通过经校准的CT室Unifors Xi CT验证的计算机断层扫描(CT)设备获得。结果显示,无论年龄组如何,ESAK值中位数范围为62.6至248.1µGy。KAP值中位数范围为135.6至1612µGy cm²,而DLP值中位数范围为119.1至600 mGy cm。结果分析表明,通过良好的实践意识以及患者剂量和图像质量评估可以实现优化。