Fernandez-Soto J M, Ten J I, Sanchez R M, España M, Pifarre X, Vano E
Medical Physics Service, Hospital Clinico San Carlos, IdISSC, Madrid, Spain Radiology Department, Facultad de Medicina, Universidad Complutense, Madrid, Spain
Radiology Service, Hospital Clinico San Carlos, IdISSC, Madrid, Spain.
Radiat Prot Dosimetry. 2015 Jul;165(1-4):53-6. doi: 10.1093/rpd/ncv043. Epub 2015 Mar 22.
The purpose of this article is to present the results of connecting the interventional radiology and cardiology laboratories of five university hospitals to a unique server using an automatic patient dose registry system (Dose On Line for Interventional Radiology, DOLIR) developed in-house, and to evaluate its feasibility more than a year after its introduction. The system receives and stores demographic and dosimetric parameters included in the MPPS DICOM objects sent by the modalities to a database. A web service provides a graphical interface to analyse the information received. During 2013, the system processed 10 788 procedures (6874 cardiac, 2906 vascular and 1008 neuro interventional). The percentages of patients requiring clinical follow-up due to potential tissue reactions before and after the use of DOLIR are presented. The system allowed users to verify in real-time, if diagnostic (or interventional) reference levels are fulfilled.