Ten J I, Fernandez J M, Vaño E
Radiology Department, San Carlos University Hospital, Madrid 28040, Spain.
Radiat Prot Dosimetry. 2011 Sep;147(1-2):325-8. doi: 10.1093/rpd/ncr350. Epub 2011 Aug 9.
The purpose of this work was to develop an automatic management system to archive and analyse the major study parameters and patient doses for fluoroscopy guided procedures performed in cardiology and interventional radiology systems. The X-ray systems used for this trial have the capability to export at the end of the procedure and via e-mail the technical parameters of the study and the patient dose values. An application was developed to query and retrieve from a mail server, all study reports sent by the imaging modality and store them on a Microsoft SQL Server data base. The results from 3538 interventional study reports generated by 7 interventional systems were processed. In the case of some technical parameters and patient doses, alarms were added to receive malfunction alerts so as to immediately take appropriate corrective actions.