Suppr超能文献

使用加和变色薄膜在临床和人体模型情况下评估的剂量图方法的准确性。

Accuracy of a dose map method assessed in clinical and anthropomorphic phantom situations using Gafchromic films.

作者信息

Bordier C, Klausz R, Desponds L

机构信息

Detection and Guidance Solutions, Image Quality and Dose Center of Excellence, GE Healthcare, 283 rue de la Minière, Buc 78530, France

Detection and Guidance Solutions, Image Quality and Dose Center of Excellence, GE Healthcare, 283 rue de la Minière, Buc 78530, France.

出版信息

Radiat Prot Dosimetry. 2015 Jul;165(1-4):244-9. doi: 10.1093/rpd/ncv034. Epub 2015 Mar 22.

Abstract

A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthropomorphic phantom situation and in a clinical situation. Dose values inside different exposed areas have been compared between the film and the dose map method. The dose map results show a good visual agreement for the anthropomorphic phantom situation, and the local doses agreed within better than 15 % compared with the Gafchromic films in both situations.

摘要

一种剂量图方法已集成到通用电气X射线血管造影系统上,以指示分布在代表个体患者形状的患者体表上的局部剂量。已进行测试,通过在拟人化体模情况和临床情况下使用Gafchromic XR-RV3胶片来评估该方法的准确性。已比较了胶片和剂量图方法在不同曝光区域内的剂量值。剂量图结果显示在拟人化体模情况下视觉一致性良好,并且在两种情况下与Gafchromic胶片相比,局部剂量的一致性优于15%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验