通过使用欧洲参考框架(EUREF)图像质量标准将低能量对比增强光谱乳腺摄影图像与全场数字化乳腺摄影图像进行比较,对其进行评估。

Evaluation of low-energy contrast-enhanced spectral mammography images by comparing them to full-field digital mammography using EUREF image quality criteria.

作者信息

Lalji U C, Jeukens C R L P N, Houben I, Nelemans P J, van Engen R E, van Wylick E, Beets-Tan R G H, Wildberger J E, Paulis L E, Lobbes M B I

机构信息

Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, P.O. Box 5800, 6202 AZ, Maastricht, The Netherlands.

出版信息

Eur Radiol. 2015 Oct;25(10):2813-20. doi: 10.1007/s00330-015-3695-2. Epub 2015 Mar 27.

Abstract

OBJECTIVE

Contrast-enhanced spectral mammography (CESM) examination results in a low-energy (LE) and contrast-enhanced image. The LE appears similar to a full-field digital mammogram (FFDM). Our aim was to evaluate LE CESM image quality by comparing it to FFDM using criteria defined by the European Reference Organization for Quality Assured Breast Screening and Diagnostic Services (EUREF).

METHODS

A total of 147 cases with both FFDM and LE images were independently scored by two experienced radiologists using these (20) EUREF criteria. Contrast detail measurements were performed using a dedicated phantom. Differences in image quality scores, average glandular dose, and contrast detail measurements between LE and FFDM were tested for statistical significance.

RESULTS

No significant differences in image quality scores were observed between LE and FFDM images for 17 out of 20 criteria. LE scored significantly lower on one criterion regarding the sharpness of the pectoral muscle (p < 0.001), and significantly better on two criteria on the visualization of micro-calcifications (p = 0.02 and p = 0.034). Dose and contrast detail measurements did not reveal any physical explanation for these observed differences.

CONCLUSIONS

Low-energy CESM images are non-inferior to FFDM images. From this perspective FFDM can be omitted in patients with an indication for CESM.

KEY POINTS

• Low-energy CESM images are non-inferior to FFDM images. • Micro-calcifications are significantly more visible on LE CESM than on FFDM. • There is no physical explanation for this improved visibility of micro-calcifications. • There is no need for an extra FFDM when CESM is indicated.

摘要

目的

对比增强光谱乳腺摄影(CESM)检查可生成低能量(LE)和对比增强图像。低能量图像看起来类似于全场数字化乳腺摄影(FFDM)。我们的目的是通过使用欧洲质量保证乳腺筛查和诊断服务参考组织(EUREF)定义的标准,将低能量CESM图像与FFDM进行比较,以评估其图像质量。

方法

两名经验丰富的放射科医生使用这20项EUREF标准,对147例同时有FFDM和低能量图像的病例进行独立评分。使用专用体模进行对比细节测量。对低能量图像和FFDM图像在图像质量评分、平均腺体剂量和对比细节测量方面的差异进行统计学显著性检验。

结果

在20项标准中的17项上,低能量图像和FFDM图像的图像质量评分没有显著差异。在一项关于胸肌清晰度的标准上,低能量图像的得分显著较低(p < 0.001),而在两项关于微钙化可视化的标准上,得分显著较高(p = 0.02和p = 0.034)。剂量和对比细节测量并未揭示这些观察到的差异的任何物理原因。

结论

低能量CESM图像不劣于FFDM图像。从这个角度来看,对于有CESM指征的患者,可以省略FFDM。

关键点

• 低能量CESM图像不劣于FFDM图像。• 低能量CESM上的微钙化比FFDM上的显著更清晰可见。• 微钙化这种改善的可见性没有物理原因。• 有CESM指征时无需额外进行FFDM检查。