Quality by design based optimization of a high performance liquid chromatography method for assay determination of low concentration preservatives in complex nasal formulations.

作者信息

Zakrajšek Jure, Stojić Vladimir, Bohanec Simona, Urleb Uroš

出版信息

Acta Chim Slov. 2015;62(1):72-82. doi: 10.17344/acsi.2014.718.

Abstract

The effects of seven different chromatographic parameters and five sample preparation parameters in a high performance liquid chromatography (HPLC) method for assay determination of benzalkonium chloride (BKC) in a nasal formulation were evaluated using two fractional factorial experimental designs. The design space of the analytical method was modeled using Umetrics Modde software and the optimal method conditions were predicted. The optimum HPLC chromatographic conditions were obtained using a Luna CN column (150 x 4.6 mm, 3 µm). The results show that mobile phase pH, amount of acetonitrile in the mobile phase and column temperature are the most important factors in obtaining good separation of BKC homologs from an interfering peak. In the sample preparation step, the use of an aqueous solution for dissolving the samples was the most important factor since it eliminated the interfering effect of the active compound. The optimal method was validated for linearity, accuracy and precision. The use of experimental designs enables obtaining the maximum amount of information with the least possible number of experiments. Such designs are an economical manner in evaluating a variety of different factors and their interactions.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验