Indication-based national diagnostic reference levels (DRLs) for a few most common paediatric computed tomography (CT) examinations are proposed. Patient dose data (CTDI vol and dose length product) were collected for over 1000 patients in 4 university hospitals with best experiences in paediatric CT. Four indications for chest CT and two for abdomen (abdomen + pelvis), chest + abdomen and head CT were considered. The DRLs for the body examinations are proposed as exponential DRL-curves, where CTDI vol and dose length product are presented as a function of patient weight. The same DRL curve applies to all the indications studied. The basic 75 % level curve is supplemented by 50 % level curve to enable considerations on varying levels of technology. For head CT, DRLs are proposed for a few age groups (1, 1-5, 5-10 and 10-15 y), separately for routine CT and CT for ventricular size. The proposed DRLs are generally lower than the few published DRLs in other countries.
本文提出了针对几种最常见儿科计算机断层扫描(CT)检查基于适应症的国家诊断参考水平(DRL)。在4家具有儿科CT最佳经验的大学医院中,收集了1000多名患者的患者剂量数据(容积CT剂量指数和剂量长度乘积)。考虑了胸部CT的四种适应症以及腹部(腹部+骨盆)、胸部+腹部和头部CT的两种适应症。身体检查的DRL以指数DRL曲线的形式提出,其中容积CT剂量指数和剂量长度乘积表示为患者体重的函数。相同的DRL曲线适用于所有研究的适应症。基本的75%水平曲线由50%水平曲线补充,以便能够考虑不同水平的技术。对于头部CT,针对几个年龄组(1岁、1 - 5岁、5 - 10岁和10 - 15岁)分别提出了常规CT和脑室大小CT的DRL。所提出的DRL通常低于其他国家公布的少数DRL。