Petersson Sven, Dyverfeldt Petter, Sigfridsson Andreas, Lantz Jonas, Carlhäll Carl-Johan, Ebbers Tino
Division of Cardiovascular Medicine, Department of Medical and Health Sciences, Linköping University, Linköping, Sweden.
Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden.
Magn Reson Med. 2016 Mar;75(3):1249-55. doi: 10.1002/mrm.25698. Epub 2015 Apr 4.
Evaluate spiral three-dimensional (3D) phase contrast MRI for the assessment of turbulence and velocity in stenotic flow.
A-stack-of-spirals 3D phase contrast MRI sequence was evaluated in vitro against a conventional Cartesian sequence. Measurements were made in a flow phantom with a 75% stenosis. Both spiral and Cartesian imaging were performed using different scan orientations and flow rates. Volume flow rate, maximum velocity and turbulent kinetic energy (TKE) were computed for both methods. Moreover, the estimated TKE was compared with computational fluid dynamics (CFD) data.
There was good agreement between the turbulent kinetic energy from the spiral, Cartesian and CFD data. Flow rate and maximum velocity from the spiral data agreed well with Cartesian data. As expected, the short echo time of the spiral sequence resulted in less prominent displacement artifacts compared with the Cartesian sequence. However, both spiral and Cartesian flow rate estimates were sensitive to displacement when the flow was oblique to the encoding directions.
Spiral 3D phase contrast MRI appears favorable for the assessment of stenotic flow. The spiral sequence was more than three times faster and less sensitive to displacement artifacts when compared with a conventional Cartesian sequence.
评估螺旋三维(3D)相位对比磁共振成像用于评估狭窄血流中的湍流和速度。
在体外将螺旋堆叠3D相位对比磁共振成像序列与传统笛卡尔序列进行对比评估。在具有75%狭窄的流动模型中进行测量。使用不同的扫描方向和流速进行螺旋成像和笛卡尔成像。计算两种方法的体积流速、最大速度和湍流动能(TKE)。此外,将估计的TKE与计算流体动力学(CFD)数据进行比较。
螺旋、笛卡尔和CFD数据的湍流动能之间具有良好的一致性。螺旋数据的流速和最大速度与笛卡尔数据吻合良好。正如预期的那样,与笛卡尔序列相比,螺旋序列的短回波时间导致位移伪影不那么明显。然而,当血流与编码方向倾斜时,螺旋和笛卡尔流速估计对位移都很敏感。
螺旋3D相位对比磁共振成像似乎有利于评估狭窄血流。与传统笛卡尔序列相比,螺旋序列速度快三倍以上,对位移伪影的敏感性更低。