Moorthy M B, Rees D G, Hope P L
Cambridge Military Hospital, Aldershot.
J R Army Med Corps. 1989 Oct;135(3):131-4. doi: 10.1136/jramc-135-03-05.
Ten preterm infants were each studied by three observers using a commercial duplex Doppler ultrasound scanner, in order to determine the intra-observer and inter-observer variability in the measurement of pulsatility index (PI) of the anterior cerebral artery. There was considerable difference in estimates of PI, with a mean range of 0.11 between the three observers for each infant. Intra-observer variability exceeded inter-observer variability and had a coefficient of variation of 8.4%. The five observers involved in this study had less than six months experience of duplex scanning. The results suggest that the PI can be measured with acceptable reproducibility by personnel with limited experience.
十名早产儿由三名观察者使用商用双功多普勒超声扫描仪进行研究,以确定大脑前动脉搏动指数(PI)测量中的观察者内和观察者间变异性。PI的估计值存在相当大的差异,每个婴儿的三名观察者之间的平均范围为0.11。观察者内变异性超过观察者间变异性,变异系数为8.4%。参与本研究的五名观察者进行双功扫描的经验不足六个月。结果表明,经验有限的人员也能以可接受的再现性测量PI。