苏丹一些医院中成人患者在使用屏片摄影和计算机X线摄影时的剂量对比研究。

A comparative study of adult patient doses in film screen and computed radiography in some Sudanese hospitals.

作者信息

Elshiekh E, Suliman I I, Habbani F

机构信息

Radiation Safety Institute, Sudan Atomic Energy Commission, PO Box 3001, Khartoum, Sudan

Radiation Safety Institute, Sudan Atomic Energy Commission, PO Box 3001, Khartoum, Sudan Department of Radiology and Molecular Imaging, Medical Physics Section; College of Medicine and Health Sciences, Sultan Qaboos University, Postcode 123, PO Box 35, Al-Khod, Oman.

出版信息

Radiat Prot Dosimetry. 2015 Jul;165(1-4):402-5. doi: 10.1093/rpd/ncv186. Epub 2015 Apr 17.

Abstract

A study was performed to compare adult patient doses in film screen (FS) and computed radiography (CR) diagnostic X-ray examinations in some hospitals in Sudan over a period of 1 y; during this period of time, the CR systems were introduced to replace FS systems. Radiation doses were estimated for 354 patients in five hospitals (two FS units and three CR units). Entrance surface air kerma (ESAK) was estimated from incident air kerma using patient exposure parameters and tube output. Dose calculations were performed using CALDOSE X 3.5 Monte Carlo-based software. In FS, third quartile of ESAK values for skull PA, skull LAT, chest PA, pelvis AP, lumbar spine AP and lumbar spine LAT were 1.5, 1.3, 0.3, 1.9, 2.8 and 5.9 mGy, respectively, while in CR, third quartile of ESAK values for the same examinations were 2.7, 1.7, 0.18, 1.7, 3.2 and 10.8 mGy, respectively. Comparable ESAK values were presented in FS and CR units. The results are important for future dose optimisation and setting national diagnostic reference levels.

摘要

进行了一项研究,以比较苏丹一些医院在1年时间里,成人患者在使用屏片(FS)和计算机X线摄影(CR)进行诊断性X线检查时所接受的剂量;在此期间,引入CR系统以取代FS系统。对五家医院(两个FS设备和三个CR设备)的354名患者的辐射剂量进行了估算。利用患者曝光参数和管输出,根据入射空气比释动能估算入射表面空气比释动能(ESAK)。使用基于蒙特卡洛的CALDOSE X 3.5软件进行剂量计算。在FS中,颅骨前后位、颅骨侧位、胸部前后位、骨盆前后位、腰椎前后位和腰椎侧位的ESAK值的第三四分位数分别为1.5、1.3、0.3、1.9、2.8和5.9 mGy,而在CR中,相同检查的ESAK值的第三四分位数分别为2.7、1.7、0.18、1.7、3.2和10.8 mGy。FS和CR设备给出了可比的ESAK值。这些结果对于未来的剂量优化和设定国家诊断参考水平具有重要意义。