Suliman I I, Abbas N, Habbani F I
Radiation Safety Institute, Sudan Atomic Energy Commission, P.O. Box 3001, Khartoum, Sudan.
Radiat Prot Dosimetry. 2007;123(2):209-14. doi: 10.1093/rpd/ncl137. Epub 2006 Sep 14.
The aim of the study was to evaluate the entrance surface doses (ESDs) to patients undergoing selected diagnostic X-ray examinations in major Sudanese hospitals. ESD per examination was estimated from X-ray tube output parameters in four hospitals comprising eight X-ray units and a sample of 346 radiographs. Hospital mean ESDs estimated range from 0.17 to 0.27 mGy for chest AP, 1.04-2.26 mGy for Skull AP/PA, 0.83-1.32 mGy for Skull LAT, 1.31-1.89 mGy for Pelvis AP, 1.46-3.33 mGy for Lumbar Spine AP and 2.9-9.9 mGy for Lumbar Spine LAT. With exception of chest PA examination at two hospitals, mean ESDs were found to be within the established international reference doses. The results are useful to national and professional organisations and can be used as a baseline upon which future dose measurements may be compared.
该研究的目的是评估苏丹主要医院中接受特定诊断性X射线检查的患者的体表入射剂量(ESD)。通过四家医院(包括八个X射线设备)的X射线管输出参数以及346张X光片样本估算每次检查的ESD。估算出的各医院平均ESD如下:胸部前后位为0.17至0.27毫戈瑞,颅骨前后位/后前位为1.04至2.26毫戈瑞,颅骨侧位为0.83至1.32毫戈瑞,骨盆前后位为1.31至1.89毫戈瑞,腰椎前后位为1.46至3.33毫戈瑞,腰椎侧位为2.9至9.9毫戈瑞。除了两家医院的胸部后前位检查外,发现平均ESD在既定的国际参考剂量范围内。这些结果对国家和专业组织很有用,可作为未来剂量测量比较的基线。