Suppr超能文献

[平面波导荧光生物传感器检测双酚A的适用性]

[Applicability of bisphenol A detection by a planar waveguide fluorescent biosensor].

作者信息

Xu Wei-cqi, Zhang Yong-ming, Zhou Xiao-hong, Shi Han-chang

出版信息

Huan Jing Ke Xue. 2015 Jan;36(1):338-42.

Abstract

As a kind of environmental endocrine disruptors, bisphenol A received a wide attention around the world. The planar waveguide fluorescent biosensor can rapidly and sensitively detect traceable bisphenol A in water samples. Under the optimized test conditions, the typical calibration curve for BPA determination by the biosensor showed a detection limit of (0.04 +/- 0.007) microg x L(-1) The linear response ranged from 0.16 microg x L(-1) to 22.40 microg x L(-1). The 50% inhibition concentration (IC50) was (1.67 +/- 0.47) microg x L(-1). By means of adding 0.5% EDTA to sample solutions could weaken the interference of water hardness. Under the optimal conditions, the recovery ratios for four kinds of real water samples were in the range of 88% to 111%, with R.S.D. less than 15%. The results indicate that this method features the potential in the practical detections of bisphenol A in real water samples.

摘要

作为一种环境内分泌干扰物,双酚A在全球受到广泛关注。平面波导荧光生物传感器能够快速、灵敏地检测水样中的痕量双酚A。在优化的测试条件下,该生物传感器测定双酚A的典型校准曲线显示检测限为(0.04±0.007)μg·L⁻¹,线性响应范围为0.16μg·L⁻¹至22.40μg·L⁻¹。50%抑制浓度(IC50)为(1.67±0.47)μg·L⁻¹。通过向样品溶液中添加0.5%的乙二胺四乙酸(EDTA)可以减弱水硬度的干扰。在最佳条件下,四种实际水样的回收率在88%至111%之间,相对标准偏差(R.S.D.)小于15%。结果表明,该方法在实际水样中双酚A的实际检测方面具有潜力。