Elsaka Shaymaa E, Elnaghy Amr M
Department of Restorative Dental Sciences, Al-Farabi College of Dentistry and Nursing.
Dent Mater J. 2015;34(3):358-63. doi: 10.4012/dmj.2014-252. Epub 2015 Apr 23.
This study compared the cyclic fatigue resistance of OneShape (OS) instrument which used continuous rotation with WaveOne (WO) instrument which used reciprocating motion. OS size 25, 0.06 taper and WO primary size 25, 0.08 taper were used in simulated canals with 45°, 60°, and 90° angles of curvature until fracture. The number of cycles to fracture (NCF) was recorded to evaluate cyclic fatigue resistance. Data were statistically analyzed using two-way analysis of variance and Tukey's HSD test. Weibull analysis was performed on NCF data. WO had a significantly higher NCF than OS for all angles of curvature (p<0.05). According to Weibull distribution, WO instrument was predicted to have a higher number of cycles to survive than OS instrument. WO instrument which used the reciprocating motion had a greater resistance to cyclic fatigue than OS which used continuous rotation movement.
本研究比较了采用连续旋转的OneShape(OS)器械与采用往复运动的WaveOne(WO)器械的抗循环疲劳性能。使用OS 25号、0.06锥度以及WO 25号、0.08锥度的器械,在模拟的45°、60°和90°弯曲角度的根管中测试直至器械折断。记录折断时的循环次数(NCF)以评估抗循环疲劳性能。数据采用双向方差分析和Tukey's HSD检验进行统计学分析。对NCF数据进行威布尔分析。在所有弯曲角度下,WO的NCF均显著高于OS(p<0.05)。根据威布尔分布,预计WO器械比OS器械具有更多的存活循环次数。采用往复运动的WO器械比采用连续旋转运动的OS器械具有更强的抗循环疲劳能力。