BMC Oral Health. 2020 Feb 4;20(1):38. doi: 10.1186/s12903-020-1027-7.
The purpose of this study was to compare the cyclic fatigue resistance of newly developed TruNatomy instruments (TRN) in single and double (S-shaped) curvature canals with HyFlex CM (HCM), Vortex Blue (VB) and RaCe (RC) instruments.
Size 20/.04 taper and size 25/0.04 of HCM, VB and RC were used. For TRN instruments, size 20/.04 taper (small) and size 26/.04 taper (prime) were used. The instruments were tested in artificial canals with double curvature (coronal curve; 60° curvature, 5 mm radius and apical curve; 70° curvature and 2 mm radius) and single curvature (60° curvature, 5 mm radius). The number of cycles to failure (NCF) was recorded. Data were statistically analyzed by Kruskal-Wallis and Dunn's multiple comparison tests. Weibull analysis was performed on NCF data. Statistical significant was set at p < 0.05.
TRN and HCM revealed higher NCF compared with the other instruments for both tested sizes in single and double curvature canals (p < 0.05). TRN and HCM showed no statistically significant difference in the NCF (p > 0.05). The probability of survival was higher for HCM and TRN instruments than VB and RC instruments.
HCM and TRN instruments were more resistant to cyclic fatigue than VB and RC instruments in single and double curvature canals. HCM and TRN instruments were anticipated to survive with higher number of cycles than the other tested instruments. RC instrument had the lowest fatigue resistance than the other instruments.
本研究旨在比较新型 TruNatomy 器械(TRN)在单弯和双弯(S 形)根管中的循环疲劳强度,与 HyFlex CM(HCM)、Vortex Blue(VB)和 RaCe(RC)器械进行比较。
使用 HCM、VB 和 RC 的 20/.04 锥度和 25/0.04 锥度,以及 TRN 的 20/.04 锥度(小号)和 26/.04 锥度(中号)。器械在具有双弯(冠向弯曲:60°曲率,5mm 半径和根尖弯曲:70°曲率,2mm 半径)和单弯(60°曲率,5mm 半径)的人工根管中进行测试。记录器械的失效循环数(NCF)。采用 Kruskal-Wallis 和 Dunn 多重比较检验对数据进行统计学分析。对 NCF 数据进行威布尔分析。p<0.05 为统计学显著差异。
TRN 和 HCM 在单弯和双弯根管中,与其他器械相比,两种尺寸的 NCF 均较高(p<0.05)。TRN 和 HCM 的 NCF 无统计学显著差异(p>0.05)。HCM 和 TRN 器械的存活率高于 VB 和 RC 器械。
在单弯和双弯根管中,HCM 和 TRN 器械比 VB 和 RC 器械更能抵抗循环疲劳。HCM 和 TRN 器械预计比其他测试器械具有更高的循环数存活率。RC 器械的疲劳强度最低。