Giner Anna, Aldaba Mikel, Arjona Montserrat, Vilaseca Meritxell, Pujol Jaume
Centre for Sensors, Instruments and Systems Development (CD6), Universitat Politècnica de Catalunya (UPC), Spain.
Centre for Sensors, Instruments and Systems Development (CD6), Universitat Politècnica de Catalunya (UPC), Spain.
Cont Lens Anterior Eye. 2015 Oct;38(5):322-6. doi: 10.1016/j.clae.2015.03.014. Epub 2015 Apr 20.
To evaluate the usefulness of an infrared open-field autorefractor as a predictor of the refractive error when fitting multifocal contact lenses (MCL).
Objective and subjective measurements of the non-cycloplegic distance refractive error were compared in patients wearing MCL. We used the Grand Seiko WAM-5500 autorefractor for the objective measurements. Three commercially available MCL were tested. Twenty-one eyes of sixteen healthy adults were included in the study. Over-refraction was evaluated in terms of spherical equivalent (SE) and astigmatic vectors (J0 and J45). The mean difference±SD of each parameter was calculated. The Kolmogorov-Smirnov test was used to verify the normal distribution. Pearson's correlation, Bland and Altman plot and paired sample t test were used to compare the results obtained with both methods.
The mean difference between objective and subjective results of the SE over-refraction was 0.13±0.42D; for astigmatic vectors J0 and J45 were 0.03±0.32D and -0.00±0.17D, respectively. The Kolmogorov-Smirnov test showed a normal distribution for all parameters. The highest Pearson's correlation coefficients were obtained for the SE with values of 0.98 without MCL and 0.97 with MCL. The lowest were obtained for J45 with values of 0.65 without MCL and 0.75 with MCL. Significant correlations were obtained for each parameter. The paired sample t test failed to show significant differences in analyzed parameters except for J0 without MCL.
The Grand Seiko WAM-5500 can be used as a screening method of over-refraction in the clinical fitting of MCL.
评估红外开放式自动验光仪在验配多焦点隐形眼镜(MCL)时作为屈光不正预测指标的有效性。
对佩戴MCL的患者进行非散瞳状态下远距离屈光不正的客观和主观测量比较。客观测量使用Grand Seiko WAM - 5500自动验光仪。测试了三种市售的MCL。16名健康成年人的21只眼纳入研究。根据等效球镜度(SE)和散光轴(J0和J45)评估过矫。计算每个参数的平均差值±标准差。采用Kolmogorov - Smirnov检验验证正态分布。使用Pearson相关性分析、Bland - Altman图和配对样本t检验比较两种方法获得的结果。
SE过矫的客观和主观结果之间的平均差值为0.13±0.42D;散光轴J0和J45的平均差值分别为0.03±0.32D和 - 0.00±0.17D。Kolmogorov - Smirnov检验显示所有参数均呈正态分布。SE的Pearson相关系数最高,未佩戴MCL时为0.98,佩戴MCL时为0.97。J45的相关系数最低,未佩戴MCL时为0.65,佩戴MCL时为0.75。各参数均有显著相关性。配对样本t检验显示,除未佩戴MCL时的J0外,各分析参数均无显著差异。
Grand Seiko WAM - 5500可作为MCL临床验配中过矫的筛查方法。