Kumar Santosh, Singh Shivanshu, Sharma Aditya Prakash, Rathi Manish
Department of Urology, Postgraduate Institute of Medical Education and Research, Chandigarh 160012, India.
Department of Nephrology, Postgraduate Institute of Medical Education and Research, Chandigarh 160012, India.
Case Rep Nephrol. 2015;2015:684976. doi: 10.1155/2015/684976. Epub 2015 Mar 30.
CAPD catheter malfunction is a common problem. Obstruction due to wrapping by appendices epiploicae of sigmoid colon has been rarely reported in literature. We report a case of CAPD catheter malfunction caused by catheter tip migration and obstruction by appendices epiploicae that was successfully managed by laparoscopic hanging loop technique. This case report highlights the ease with which epiplopexy can be performed and catheter tip migration can be prevented by this innovative laparoscopic procedure.
持续性非卧床腹膜透析(CAPD)导管故障是一个常见问题。文献中很少报道因乙状结肠网膜包裹导致的梗阻。我们报告一例CAPD导管故障,其由导管尖端移位和网膜梗阻引起,通过腹腔镜悬吊环技术成功处理。本病例报告强调了通过这种创新的腹腔镜手术进行网膜固定和预防导管尖端移位的简便性。