An improved electrostatic integrating radon monitor with the CR-39 as alpha-particle detector.

作者信息

Fan D, Zhuo W, Chen B, Zhao C, Yi Y, Zhang Y

机构信息

Institute of Radiation Medicine, Fudan University, 2094 Xietu Road, Shanghai 200032, China.

Institute of Radiation Medicine, Fudan University, 2094 Xietu Road, Shanghai 200032, China

出版信息

Radiat Prot Dosimetry. 2015 Nov;167(1-3):121-5. doi: 10.1093/rpd/ncv227. Epub 2015 Apr 27.

Abstract

In this study, based on the electrostatic integrating radon monitor (EIRM) developed by Iida et al., a new type of EIRM with the allyl glycol carbonate (CR-39) as alpha-particle detector was developed for outdoor radon measurements. Besides using the CR-39 to replace the cellulose nitrate film as alpha-particle detector, the electrode and the setting place of the CR-39 were also optimally designed based on the simulation results of the electric field and the detection efficiency. The calibration factor of the new EIRM was estimated to be 0.136±0.002 tracks cm(-2) (Bq m(-3) h)(-1), with the lower detection limit of 0.6 Bq m(-3) for a 2-month exposure. Furthermore, both the battery and the dry agent were also replaced to protect the environment. The results of intercomparison and field experiments showed that the performances of the new EIRM were much better than the original one. It suggests that the new type of ERIM is more suitable for large-scale and long-term outdoor radon surveys.