De Crop An, Casselman Jan, Van Hoof Tom, Dierens Melissa, Vereecke Elke, Bossu Nicolas, Pamplona Jaime, D'Herde Katharina, Thierens Hubert, Bacher Klaus
Ghent University, Department of Basic Medical Sciences, Proeftuinstraat 86, B-9000, Gent, Belgium,
Neuroradiology. 2015 Aug;57(8):841-9. doi: 10.1007/s00234-015-1537-1. Epub 2015 May 1.
Metal artifacts may negatively affect radiologic assessment in the oral cavity. The aim of this study was to evaluate different metal artifact reduction techniques for metal artifacts induced by dental hardware in CT scans of the oral cavity.
Clinical image quality was assessed using a Thiel-embalmed cadaver. A Catphan phantom and a polymethylmethacrylate (PMMA) phantom were used to evaluate physical-technical image quality parameters such as artifact area, artifact index (AI), and contrast detail (IQFinv). Metal cylinders were inserted in each phantom to create metal artifacts. CT images of both phantoms and the Thiel-embalmed cadaver were acquired on a multislice CT scanner using 80, 100, 120, and 140 kVp; model-based iterative reconstruction (Veo); and synthesized monochromatic keV images with and without metal artifact reduction software (MARs). Four radiologists assessed the clinical image quality, using an image criteria score (ICS).
Significant influence of increasing kVp and the use of Veo was found on clinical image quality (p = 0.007 and p = 0.014, respectively). Application of MARs resulted in a smaller artifact area (p < 0.05). However, MARs reconstructed images resulted in lower ICS.
Of all investigated techniques, Veo shows to be most promising, with a significant improvement of both the clinical and physical-technical image quality without adversely affecting contrast detail. MARs reconstruction in CT images of the oral cavity to reduce dental hardware metallic artifacts is not sufficient and may even adversely influence the image quality.
金属伪影可能会对口腔的放射学评估产生负面影响。本研究的目的是评估不同的金属伪影减少技术,用于口腔CT扫描中由牙科硬件引起的金属伪影。
使用蒂尔(Thiel)防腐尸体评估临床图像质量。使用Catphan体模和聚甲基丙烯酸甲酯(PMMA)体模来评估物理技术图像质量参数,如伪影面积、伪影指数(AI)和对比细节(IQFinv)。在每个体模中插入金属圆柱体以产生金属伪影。使用80、100、120和140 kVp在多层CT扫描仪上采集两个体模和蒂尔防腐尸体的CT图像;基于模型的迭代重建(Veo);以及使用和不使用金属伪影减少软件(MARs)的合成单色keV图像。四名放射科医生使用图像标准评分(ICS)评估临床图像质量。
发现增加kVp和使用Veo对临床图像质量有显著影响(分别为p = 0.007和p = 0.014)。应用MARs导致伪影面积较小(p < 0.05)。然而,MARs重建图像导致ICS较低。
在所有研究的技术中,Veo显示出最有前景,在不影响对比细节的情况下,临床和物理技术图像质量均有显著改善。在口腔CT图像中使用MARs重建来减少牙科硬件金属伪影是不够的,甚至可能对图像质量产生不利影响。