Ji Xuanrong, Xiong Kedi, Yang Sihua, Xing Da
Opt Express. 2015 Apr 6;23(7):9130-6. doi: 10.1364/OE.23.009130.
We have developed an intravascular confocal photoacoustic (PA) endoscope with symmetrically aligned dual-element ultrasonic transducers. By combining focused laser excitation and focused acoustic collection, the intravascular confocal PA endoscope is capable of realizing resolution enhanced intravascular PA imaging with improved signal-to-noise ratio (SNR) to ameliorate the resolution reduction caused by laser scattering with increasing tissue depth. The detection sensitivity of the endoscope is improved by 5 dB compared with that of single transducer endoscope, and the transverse resolution of the system can up to 13 μm. Intravascular PA tomography of a normal vessel and an atherosclerotic vessel have been performed to demonstrate the imaging ability of the system. This intravascular confocal PA endoscope with an outer diameter of 1.2 mm supports potential for clinical applications in intravascular plaque imaging and subsequent diagnosis.
我们开发了一种具有对称排列双元件超声换能器的血管内共聚焦光声(PA)内窥镜。通过结合聚焦激光激发和聚焦声采集,血管内共聚焦PA内窥镜能够实现分辨率增强的血管内PA成像,同时提高信噪比(SNR),以改善随着组织深度增加因激光散射导致的分辨率降低问题。与单换能器内窥镜相比,该内窥镜的检测灵敏度提高了5 dB,系统的横向分辨率可达13μm。已对正常血管和动脉粥样硬化血管进行了血管内PA断层扫描,以证明该系统的成像能力。这种外径为1.2 mm的血管内共聚焦PA内窥镜在血管内斑块成像及后续诊断的临床应用方面具有潜力。