Schanzer H, Martinelli G, Burrows L, Chiang K, Peirce E C
Department of Surgery, Mount Sinai School of Medicine New York, New York 10029.
ASAIO Trans. 1989 Jul-Sep;35(3):211-3. doi: 10.1097/00002480-198907000-00011.
A double wall PTFE graft with a silicone rubber middle layer (PTFE-sil) shown in a canine model to be self-sealing after needle puncture, was tried in 30 dialysis patients. Thirty-five patients with PTFE grafts were controls. All patients were followed for 1 year. In the PTFE-sil group, mean time to first dialysis was 1.3 days after implantation; during first 10 treatments, 17% of punctures did not bleed, 61% bled minimally, and 4% required 10 or more minutes of compression. No bleeding complications were encountered at any time. Complications of PTFE-sil vs. PTFE were as follows: thrombosis 36.6% vs. 28.5%; infection 13.3% vs. 11.4%; seroma 3% vs. 0%; steal 3% vs. 0%; pseudoaneurysm 0% vs. 17.1%; one year primary patency 63% vs. 66%; secondary patency 75% vs. 67%. In conclusion, the PTFE-sil graft, can be used immediately after implantation, sparing patient additional access procedures for acute dialysis. This results in less morbidity, decrease in bleeding complications, considerable shortening of hospital stay and significant reduction in expenses.
一种带有硅橡胶中间层的双层聚四氟乙烯移植物(PTFE-sil)在犬类模型中显示针刺后能自我封闭,在30例透析患者中进行了试验。35例使用聚四氟乙烯移植物的患者作为对照。所有患者均随访1年。在PTFE-sil组,首次透析的平均时间为植入后1.3天;在前10次治疗中,17%的穿刺点不出血,61%轻微出血,4%需要压迫10分钟或更长时间。在任何时候均未遇到出血并发症。PTFE-sil与聚四氟乙烯的并发症如下:血栓形成36.6%对28.5%;感染13.3%对11.4%;血清肿3%对0%;窃血3%对0%;假性动脉瘤0%对17.1%;一年原发性通畅率63%对66%;继发性通畅率75%对67%。总之,PTFE-sil移植物植入后可立即使用,避免了患者进行急性透析的额外通路操作。这导致发病率降低、出血并发症减少、住院时间显著缩短和费用大幅降低。