Cerny R, Johnova K, Kozlovska M, Otahal P, Vosahlikova I
National Institute for NBC Protection, Kamenna 71, Milin 262 31, Czech Republic CTU in Prague, FNSPE, DDAIR, Brehova 7, Praha 1 115 19, Czech Republic
CTU in Prague, FNSPE, DDAIR, Brehova 7, Praha 1 115 19, Czech Republic.
Radiat Prot Dosimetry. 2015 Jun;164(4):533-6. doi: 10.1093/rpd/ncv338. Epub 2015 May 15.
For experiments with dispersed radioactive aerosols in a radon-aerosol chamber (RAC), it is desirable to know the activity of the radioactive aerosols applied in the RAC. A COLIBRI TTC survey metre with an SABG-15+ probe (Canberra, USA) was purchased for this purpose. The probe is designed for surface contamination measurements, and it is intended to measure the activity of aerosols deposited on the filters during experiments in the RAC. Since the probe is calibrated in a different geometry, its response in the authors' experimental geometry was simulated by a Monte Carlo method. The authors present a Monte Carlo model using MCNPX and an experimental verification of this probe model.