Suppr超能文献

难治性青光眼患者的高强度聚焦超声治疗:前瞻性临床研究1年结果

High-intensity focused ultrasound treatment in refractory glaucoma patients: results at 1 year of prospective clinical study.

作者信息

Melamed Shlomo, Goldenfeld Modi, Cotlear Daniel, Skaat Alon, Moroz Iris

机构信息

The Sam Rothberg Glaucoma Center, Goldschleger Eye Institute, Sheba Medical Center, Tel Aviv University, Tel Hashomer - Israel.

出版信息

Eur J Ophthalmol. 2015 Nov-Dec;25(6):483-9. doi: 10.5301/ejo.5000620. Epub 2015 May 13.

Abstract

PURPOSE

To evaluate the safety and efficacy of the ultrasonic circular cyclo-coagulation procedure using high-intensity focused ultrasound by a miniaturized annular device containing 6 piezoceramic transducers in patients with refractory glaucoma.

METHODS

This was a prospective interventional noncomparative study of 20 eyes of 20 patients with refractory glaucoma. All eyes were treated with 6 activated transducers operating at 21 MHz. Ultrasound biomicroscopy and a complete ophthalmic examination were performed before the procedure and at 1 day, 1 week, and 1, 3, 6, and 12 months after the procedure. Primary outcomes were surgical success (defined as intraocular pressure (IOP) reduction from baseline ≥20% and IOP >5 mm Hg) at the last follow-up visit. Secondary outcomes were mean IOP at each follow-up visit compared to baseline, medication use, complications, and re-interventions.

RESULTS

Intraocular pressure was significantly reduced (p<0.01) from a mean preoperative value of 36.4 ± 5.7 mm Hg to a mean postoperative value of 22.5 ± 10.3 mm Hg at 12 months. Four patients needed to be re-treated. The mean IOP reduction achieved was 38%. Surgical success was achieved in 13 of 20 eyes (65%). No major intraoperative or postoperative complications occurred.

CONCLUSIONS

Ultrasonic circular cyclo-coagulation using high-intensity focused ultrasound delivered by a circular miniaturized device containing 6 piezoceramic transducers is an effective and well-tolerated method to reduce IOP in patients with refractory glaucoma.

摘要

目的

评估使用包含6个压电陶瓷换能器的小型环形装置,通过高强度聚焦超声进行超声环形睫状体凝固术治疗难治性青光眼患者的安全性和有效性。

方法

这是一项针对20例难治性青光眼患者的20只眼睛的前瞻性介入性非对照研究。所有眼睛均使用6个在21 MHz频率下工作的激活换能器进行治疗。在手术前以及手术后1天、1周、1个月、3个月、6个月和12个月进行超声生物显微镜检查和全面的眼科检查。主要结局是在最后一次随访时手术成功(定义为眼压从基线降低≥20%且眼压>5 mmHg)。次要结局是每次随访时与基线相比的平均眼压、药物使用情况、并发症和再次干预情况。

结果

眼压从术前平均36.4±5.7 mmHg显著降低(p<0.01)至术后12个月时的平均22.5±10.3 mmHg。4例患者需要再次治疗。平均眼压降低幅度为38%。20只眼中有13只(65%)手术成功。未发生重大术中或术后并发症。

结论

使用包含6个压电陶瓷换能器的圆形小型装置进行高强度聚焦超声的超声环形睫状体凝固术,是降低难治性青光眼患者眼压的一种有效且耐受性良好的方法。