Abstract

PURPOSE

The aim of this study was to evaluate the safety and efficacy of ultrasound cyclo-plasty (UCP) for reducing intraocular pressure (IOP) in patients with glaucoma.

METHODS

This is a multicentre prospective study conducted in 3 Italian glaucoma centres. UCP was performed by EyeOP1, which delivers ultrasound beams using 6 piezoelectric transducers activated for 4/6 s (first generation) or 8 s (second generation). Primary outcomes were the mean IOP reduction and the rates of success after 1 year. Secondary outcomes were the mean IOP reduction at each follow-up, and the reduction of the number of hypotensive medications.

RESULTS

In total, 49 eyes from 47 patients were treated. One year postoperatively, the mean IOP had decreased from 27.7 ± 9.2 to 19.8 ± 6.9 mm Hg (p < 0.001), and the mean number of hypotensive drops and tablets had decreased from 3.2 and 0.5 to 2.3 and 0.2, respectively (p < 0.05). Postoperative IOP reduction was significantly related to preoperative IOP (r2 = 0.5034; p < 0.0001). Second-generation probes determined a significantly higher IOP reduction (p < 0.05). Qualified success was achieved in 25 eyes (51.1%) and complete success in 21 (42.9%), while failure was recorded in 12 (24.5%).

CONCLUSIONS

UCP is safe and effective for reducing IOP. The procedure determines a greater IOP reduction in patients with higher preoperative IOP. Second-generation probes improve outcomes without detrimental effects on safety.

摘要

目的

本研究旨在评估超声睫状体成形术(UCP)降低青光眼患者眼压(IOP)的安全性和有效性。

方法

这是一项在3个意大利青光眼中心进行的多中心前瞻性研究。UCP由EyeOP1设备实施,该设备使用6个压电换能器发射超声束,第一代激活4/6秒,第二代激活8秒。主要结局指标为1年后平均眼压降低幅度和成功率。次要结局指标为每次随访时的平均眼压降低幅度以及降压药物数量的减少。

结果

共治疗了47例患者的49只眼。术后1年,平均眼压从27.7±9.2毫米汞柱降至19.8±6.9毫米汞柱(p<0.001),降压滴眼液和片剂的平均数量分别从3.2和0.5降至2.3和0.2(p<0.05)。术后眼压降低与术前眼压显著相关(r2 = 0.5034;p<0.0001)。第二代探头使眼压降低幅度显著更高(p<0.05)。25只眼(51.1%)取得合格成功,21只眼(42.9%)取得完全成功,12只眼(24.5%)记录为失败。

结论

UCP降低眼压安全有效。该手术使术前眼压较高的患者眼压降低幅度更大。第二代探头改善了治疗效果且对安全性无不利影响。