Singh S, Schaaf N G
Int J Oral Maxillofac Implants. 1989 Summer;4(2):139-46.
All implantable devices must be sterile. However, autoclaves produce poor surface properties that jeopardize the integration process. The application of a modified ultraviolet light source has proven to enhance bioreactivity by controlling surface properties, but it lacks validation of its sterilization capabilities. Forty-eight titanium implants were contaminated with spores of the biological indicator Bacillus stearothermophilus and subjected to "dynamic sterilization" by ultraviolet light. Forty-seven of the implants were successfully sterilized, as indicated by not producing turbidity in a suitable growth medium. This sterilization technique only requires a 20-second exposure to achieve sterility.