Haubenreisser Holger, Meyer Mathias, Sudarski Sonja, Allmendinger Thomas, Schoenberg Stefan O, Henzler Thomas
Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University, Germany.
Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University, Germany.
Eur J Radiol. 2015 Aug;84(8):1608-1613. doi: 10.1016/j.ejrad.2015.04.018. Epub 2015 Apr 28.
To prospectively investigate image quality and radiation dose of 100kVp spectral shaping chest CT using a dedicated tin filter on a 3rd generation dual-source CT (DSCT) in comparison to standard 100kVp chest CT.
Sixty patients referred for a non-contrast chest on a 3rd generation DSCT were prospectively included and examined at 100kVp with a dedicated tin filter. These patients were retrospectively matched with patients that were examined on a 2nd generation DSCT at 100kVp without tin filter. Objective and subjective image quality was assessed in various anatomic regions and radiation dose was compared.
Radiation dose was decreased by 90% using the tin filter (3.0 vs 0.32mSv). Soft tissue attenuation and image noise was not statistically different for both examination techniques (p>0.05), however image noise was found to be significantly higher in the trachea when using the additional tin filter (p=0.002). SNR was found to be statistically similar in pulmonary tissue, significantly lower when measured in air and significantly higher in the aorta for the scans on the 3rd generation DSCT. Subjective image quality with regard to overall quality and image noise and sharpness was not statistically significantly different (p>0.05).
100kVp spectral shaping chest CT by means of a tube-based tin-filter on a 3rd generation DSCT allows 90% dose reduction when compared to 100kVp chest CT on a 2nd generation DSCT without spectral shaping.
前瞻性研究在第三代双源CT(DSCT)上使用专用锡滤过器的100kVp光谱塑形胸部CT的图像质量和辐射剂量,并与标准100kVp胸部CT进行比较。
前瞻性纳入60例因非增强胸部检查而行第三代DSCT检查的患者,使用专用锡滤过器在100kVp下进行检查。这些患者与在第二代DSCT上100kVp未使用锡滤过器检查的患者进行回顾性匹配。评估不同解剖区域的客观和主观图像质量,并比较辐射剂量。
使用锡滤过器时辐射剂量降低了90%(3.0 vs 0.32mSv)。两种检查技术的软组织衰减和图像噪声无统计学差异(p>0.05),然而,使用额外锡滤过器时气管内的图像噪声明显更高(p=0.002)。第三代DSCT扫描的肺组织信噪比在统计学上相似,在空气中测量时显著降低,在主动脉中显著升高。在整体质量、图像噪声和清晰度方面的主观图像质量无统计学显著差异(p>0.05)。
与未进行光谱塑形的第二代DSCT上的100kVp胸部CT相比,在第三代DSCT上通过基于管的锡滤过器进行100kVp光谱塑形胸部CT可使剂量降低90%。