Bodelle Boris, Fischbach Constanze, Booz Christian, Yel Ibrahim, Frellesen Claudia, Kaup Moritz, Beeres Martin, Vogl Thomas J, Scholtz Jan-Erik
Department of Diagnostic and Interventional Radiology, Goethe University of Frankfurt, Theodor-Stern-Kai 7, 60590, Frankfurt, Germany.
Pediatr Radiol. 2017 Jun;47(7):831-837. doi: 10.1007/s00247-017-3813-1. Epub 2017 Mar 28.
Most of the applied radiation dose at CT is in the lower photon energy range, which is of limited diagnostic importance.
To investigate image quality and effects on radiation parameters of 100-kVp spectral filtration single-energy chest CT using a tin-filter at third-generation dual-source CT in comparison to standard 100-kVp chest CT.
Thirty-three children referred for a non-contrast chest CT performed on a third-generation dual-source CT scanner were examined at 100 kVp with a dedicated tin filter with a tube current-time product resulting in standard protocol dose. We compared resulting images with images from children examined using standard single-source chest CT at 100 kVp. We assessed objective and subjective image quality and compared radiation dose parameters.
Radiation dose was comparable for children 5 years old and younger, and it was moderately decreased for older children when using spectral filtration (P=0.006). Effective tube current increased significantly (P=0.0001) with spectral filtration, up to a factor of 10. Signal-to-noise ratio and image noise were similar for both examination techniques (P≥0.06). Subjective image quality showed no significant differences (P≥0.2).
Using 100-kVp spectral filtration chest CT in children by means of a tube-based tin-filter on a third-generation dual-source CT scanner increases effective tube current up to a factor of 10 to provide similar image quality at equivalent dose compared to standard single-source CT without spectral filtration.
CT检查中大部分应用的辐射剂量处于较低光子能量范围,其诊断价值有限。
与标准100 kVp胸部CT相比,研究在第三代双源CT上使用锡滤过进行100 kVp光谱滤过单能量胸部CT的图像质量及对辐射参数的影响。
对33例因非增强胸部CT检查而转诊至第三代双源CT扫描仪的儿童,在100 kVp下使用专用锡滤过进行检查,管电流-时间乘积达到标准方案剂量。我们将所得图像与100 kVp下使用标准单源胸部CT检查的儿童图像进行比较。我们评估了客观和主观图像质量,并比较了辐射剂量参数。
5岁及以下儿童的辐射剂量相当,而对于年龄较大的儿童,使用光谱滤过时辐射剂量适度降低(P = 0.006)。光谱滤过使有效管电流显著增加(P = 0.0001),增加幅度高达10倍。两种检查技术的信噪比和图像噪声相似(P≥0.06)。主观图像质量无显著差异(P≥0.2)。
在第三代双源CT扫描仪上通过基于管的锡滤过对儿童使用100 kVp光谱滤过胸部CT,可使有效管电流增加高达10倍,从而在等效剂量下提供与无光谱滤过的标准单源CT相似的图像质量。