Urolithiasis. 2023 Mar 18;51(1):54. doi: 10.1007/s00240-023-01419-5.
To investigate the optimal scanning parameters of dual-energy computed tomography (DECT), which can accurately determine sensitivity (the detectability of urinary stones) and accuracy (the composition matching of urinary stones), and to apply them to clinical trials. Fifteen urinary stones were chemically analyzed, and their chemical compositions were considered a reference standard with which we compared the uric acid (UA) and non-UA compositions determined using DECT. The urinary stones were placed inside a bolus and scanned with a dual-source CT scanner under various selected dual-energy conditions (A to X) using various solid water phantom thicknesses. These datasets were analyzed using the Siemens syngo.via software tool (integrated into the CT system) for matching the sensitivity and accuracy assessments. This study showed that 80% of the highest sensitivity (detection of urinary stones) and 92% of the highest accuracy (composition matching of urinary stones) were achieved under condition A (a collimation beam width setting of 2 × 32 × 0.6 mm, an automatic exposure control setting of 80/sn140 peak kilovoltage, and a slice thickness of 0.5/0.5 mm) (P < 0.05). Application of the DECT energy parameters presented in the study will help identify the sensitivity and accuracy of UA and non-UA stone analysis, even in patients with small-sized urinary stones and in conditions difficult for analysis.
为了研究双能 CT(DECT)的最佳扫描参数,以准确确定灵敏度(尿路结石的可检测性)和准确性(尿路结石的成分匹配),并将其应用于临床试验。我们对 15 颗尿路结石进行了化学分析,将其化学成分作为参考标准,与 DECT 确定的尿酸(UA)和非-UA 成分进行比较。将这些尿路结石置于球囊内,并在各种选定的双能条件(A 至 X)下使用各种实心水体模厚度,使用双源 CT 扫描仪进行扫描。使用西门子 syngo.via 软件工具(集成在 CT 系统中)对这些数据集进行分析,以进行灵敏度和准确性评估。本研究表明,在条件 A(准直器宽度设置为 2×32×0.6mm、自动曝光控制设置为 80/sn140 峰值千伏、层厚为 0.5/0.5mm)下,灵敏度(检测尿路结石)最高可达 80%,准确性(尿路结石成分匹配)最高可达 92%(P<0.05)。应用本研究中提出的 DECT 能量参数将有助于确定 UA 和非-UA 结石分析的灵敏度和准确性,即使在结石较小和分析困难的患者中也是如此。